AEROSET/ARCHITECT c8000 Urea Nitrogen Models: 7D75-01; 7D75-20; 7D75-30
Recall: Z-0240-04
·
Initiated November 24, 2003
Recall
- Recall Number
- Z-0240-04
- Event Number
- 27822
- Firm
- Abbott Laboratories Diagnostic Div
- FEI Number
- 2018433
- Product Code
- CDQ
- Status
- Terminated
- Root Cause
- Other
- Initiated
- November 24, 2003
- Posted
- December 18, 2003
- Terminated
- January 25, 2012
- Address
- 820 Mission St, South Pasadena, CA, 91030-3142
Description
AEROSET/ARCHITECT c8000 Urea Nitrogen Models: 7D75-01; 7D75-20; 7D75-30
Reason
False low results may occur with out error codes due to substrate depletion in samples with more than 220mg/dL BUN.
Action
Firm sent letter to customers dated November 24,2003 which includes a new package insert to use with this reagent. Instructions were changed to correct low resulting situation. Subrecall was requested.
Distribution
Nationwide and Mexico, Brazil, Germany, Singapore, Grand Cayman, Venezuela, Puerto Rico, Hong Kong, Australia, Columbia, Dominican Republic, China, Trinidad, Argentina, Canada, New Zealand, Bahamas
Quantity
All lots.