FDA Recall Terminated

UniCel DxH Slidemaker Stainer, Part Number 75222, for use with automated hematology analyzer

Recall: Z-0238-2018 · Initiated November 3, 2017

Recall

Recall Number
Z-0238-2018
Event Number
78611
Firm
Beckman Coulter Inc.
FEI Number
2050012
Product Code
GKZ
Status
Terminated
Root Cause
Device Design
Initiated
November 3, 2017
Terminated
June 19, 2018
Address
250 S Kraemer Blvd, Brea, CA, 92821-6232

Description

UniCel DxH Slidemaker Stainer, Part Number 75222, for use with automated hematology analyzer

Reason

The shield for the Single-Tube Station on the DxH instrument may interfere with sample aspiration in the open vial position.

Action

The firm notified its consignees of the recall on 11/03/2017 by letter. The letter stated the issue and provided the following instructions: "To avoid a biohazardous spill, do not use the right (green) position of the Single-Tube Station for sample aspiration until the shield is removed. You may remove the shield following the instructions in Attachment 1 or contact your local Beckman Coulter Representative for further assistance."

Distribution

CA, AR, HI, CA, OH, WA, MD, Canada, Japan

Quantity

3 units