FDA Recall Open, Classified

Change Healthcare Radiology Solutions, software versions 14.0, 14.1, and 14.2 Medical image and information management software.

Recall: Z-0237-2024 · Initiated October 12, 2023

Recall

Recall Number
Z-0237-2024
Event Number
93259
Firm
CHANGE HEALTHCARE CANADA COMPANY
FEI Number
3003757840
Product Code
LLZ
Status
Open, Classified
Root Cause
Software design
Initiated
October 12, 2023
Posted
November 6, 2023
Address
10711 Cambie Rd, Suite 130, Richmond Canada

Description

Change Healthcare Radiology Solutions, software versions 14.0, 14.1, and 14.2 Medical image and information management software.

Reason

Change Healthcare has identified a software issue where Change Healthcare Radiology Solutions 14.0, 14.1, and 14.2 may not update records in external applications.

Action

Field Safety Notice was distributed to all consignees on10/12/2023 by email. The notice explained the issue, potential risk to the patient, and directed users to to check the CHRS Exception Tracker on a regular basis, as failed updates will be listed with details on the failure. Recommended actions to take on potentially affected data: Impacted data can be identified by using the CHRS Exception Tracker. Customers are advised to review records of recent updates from CHRS to external systems and manually update any changes to study records that may have been missed. Change Healthcare will install a software update.

Distribution

US, Canada, Ireland, Saudi Arabia, France, UK

Quantity

282 units