UniCel DxH600, Part Number B23858, automated hematology analyzer (Note: the DxH 600 is the same instrument as the DxH 800 except that it is a bench top unit with no instrument cart, computer stand, and no connectivity capabilities)
Recall
- Recall Number
- Z-0237-2018
- Event Number
- 78611
- Firm
- Beckman Coulter Inc.
- FEI Number
- 2050012
- Product Code
- GKZ
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- November 3, 2017
- Terminated
- June 19, 2018
- Address
- 250 S Kraemer Blvd, Brea, CA, 92821-6232
Description
UniCel DxH600, Part Number B23858, automated hematology analyzer (Note: the DxH 600 is the same instrument as the DxH 800 except that it is a bench top unit with no instrument cart, computer stand, and no connectivity capabilities)
The shield for the Single-Tube Station on the DxH instrument may interfere with sample aspiration in the open vial position.
The firm notified its consignees of the recall on 11/03/2017 by letter. The letter stated the issue and provided the following instructions: "To avoid a biohazardous spill, do not use the right (green) position of the Single-Tube Station for sample aspiration until the shield is removed. You may remove the shield following the instructions in Attachment 1 or contact your local Beckman Coulter Representative for further assistance."
CA, AR, HI, CA, OH, WA, MD, Canada, Japan
1 unit