FDA Recall Terminated

UniCel DxH600, Part Number B23858, automated hematology analyzer (Note: the DxH 600 is the same instrument as the DxH 800 except that it is a bench top unit with no instrument cart, computer stand, and no connectivity capabilities)

Recall: Z-0237-2018 · Initiated November 3, 2017

Recall

Recall Number
Z-0237-2018
Event Number
78611
Firm
Beckman Coulter Inc.
FEI Number
2050012
Product Code
GKZ
Status
Terminated
Root Cause
Device Design
Initiated
November 3, 2017
Terminated
June 19, 2018
Address
250 S Kraemer Blvd, Brea, CA, 92821-6232

Description

UniCel DxH600, Part Number B23858, automated hematology analyzer (Note: the DxH 600 is the same instrument as the DxH 800 except that it is a bench top unit with no instrument cart, computer stand, and no connectivity capabilities)

Reason

The shield for the Single-Tube Station on the DxH instrument may interfere with sample aspiration in the open vial position.

Action

The firm notified its consignees of the recall on 11/03/2017 by letter. The letter stated the issue and provided the following instructions: "To avoid a biohazardous spill, do not use the right (green) position of the Single-Tube Station for sample aspiration until the shield is removed. You may remove the shield following the instructions in Attachment 1 or contact your local Beckman Coulter Representative for further assistance."

Distribution

CA, AR, HI, CA, OH, WA, MD, Canada, Japan

Quantity

1 unit