FDA Recall Terminated

Surgical packs (includes trays, bags and kits) Used by medical staff to perform patient care procedures such as wound healing, suture removals, irrigate wounds, etc.

Recall: Z-0237-2015 · Initiated October 8, 2014

Recall

Recall Number
Z-0237-2015
Event Number
69448
Firm
Customed, Inc Calle Igualdad Final # 7
FEI Number
2648727
Product Code
MCZ
Status
Terminated
Root Cause
Packaging
Initiated
October 8, 2014
Posted
December 16, 2014
Terminated
August 14, 2017
Address
Fajardo, PR, 00738

Description

Surgical packs (includes trays, bags and kits) Used by medical staff to perform patient care procedures such as wound healing, suture removals, irrigate wounds, etc.

Reason

Customed products are being recalled because of compromised sterility, not limited to potentially damaged packaging. Serious deficiencies in the manufacturing processes (QSR) and uncontrolled and inadequate storage conditions increase the risk of contamination and may result in patient infection.

Action

Customed initiated a recall on October 8th, 2014 of sterile convenience surgical Packs-trays, bags (known as Convenience Packs) and new lots of products added to the initial June 3, 2014 recall of 233 sterile surgical packs. The products are being recalled because of significant compromised sterility, not limited to potentially damaged packaging. Customers were instructed to quarantine and cease use of the affected product identified. Customers are to return the response form included with the letter, to the contact information provided. Upon receipt, the firm will contact customers with further instructions to obtain credit, replacement product, and for return of the recalled product or any other disposition action. If customers have any questions, they should contact the Customed Recall Coordinator at 407-850-5558 extensions 8001 or 8005, or via e-mail: [email protected].

Distribution

US Distribution to Florida, New York and Puerto Rico..

Quantity

185000