3M Attest 1264P Biological Indicators, in boxes of 25 per box

Recall: Z-0237-04 · Initiated October 24, 2003

Recall

Recall Number
Z-0237-04
Event Number
27465
FEI Number
2110898
Product Code
FRC
Status
Terminated
Root Cause
Other
Initiated
October 24, 2003
Posted
December 18, 2003
Terminated
February 9, 2005
Address
3m Health Care 3m Center, Saint Paul, MN, 55144-1001

Description

3M Attest 1264P Biological Indicators, in boxes of 25 per box

Reason

The 3M Attest 1264/1264P Biological Indicators of ETO sterilization contain a microbiological contaminant which can affect the performance of the positive control and the indicators to some limited extent. The color on a positive control may revert to negative after 24 hours of incubation.

Action

The firm conducted a field correction which began on 10/24/03. They provided revised labeling for the product. Letters, dated October 24, 2003, were sent to consignees. The letters had a description of the problem, instructions for doing the relabeling in the field, and recommendations for users of the product.

Distribution

The product was distributed nationwide in the United States and worldwide.

Quantity

1844 boxes of 25 biological indicators per box