Stryker Renaissance Series/Advantage Series PACU wheeled stretcher, Model 1509, Stryker Medical, Portage, MI 49002
Recall
- Recall Number
- Z-0236-2008
- Event Number
- 37214
- Firm
- Stryker Medical Div. of Stryker Corporation
- FEI Number
- 1831750
- Product Code
- FPO
- Status
- Terminated
- Root Cause
- Component change control
- Initiated
- November 15, 2004
- Posted
- November 10, 2007
- Terminated
- October 27, 2010
- Address
- 3800 E. Centre Ave., Portage, MI, 49002
Description
Stryker Renaissance Series/Advantage Series PACU wheeled stretcher, Model 1509, Stryker Medical, Portage, MI 49002
The brake cams may crack or develop excessive wear, resulting in the brakes not locking the wheels, or the stretcher not being locked firmly in position.
The firm issued an Urgent Medical Device Correction Letter dated 9/24/07, to each consignee informing them of the problems, that the brakes on each affected stretcher needed to be inspected/tested and removed from service if brake problems were discovered, and that a Stryker service representative would be in contact to replace the brakes.
Worldwide-USA, Argentina, Australia, Brazil, Canada, Chile, China, Dubai, France, India, Italy, Japan, Germany, Greece, Hong Kong, Korea, Latin America, Malaysia, Mexico, Netherlands, Poland, Portugal, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, and United Kingdom.
81,367 (61,708 U.S. and 19,569 ex-U.S.) for all models.