FDA Recall Open, Classified

Atrium Ocean Wet Suction Water Seal Chest Drain- Product Code: 2002-000

Recall: Z-0235-2023 · Initiated October 18, 2022

Recall

Recall Number
Z-0235-2023
Event Number
91043
Firm
Atrium Medical Corporation
FEI Number
3011175548
Product Code
KDQ
Status
Open, Classified
Root Cause
Storage
Initiated
October 18, 2022
Posted
November 15, 2022
Address
40 Continental Blvd, Merrimack, NH, 03054-4332

Description

Atrium Ocean Wet Suction Water Seal Chest Drain- Product Code: 2002-000

Reason

Warehouse experienced temperature excursions in July and August 2021. It was determined that there were no associated harms to patients.

Action

Atrium/ Getinge issued Urgent Medical Device Removal on 10/18/21 via email with Delivery and Read receipt. Letter states reason for recall, health risk and action to take: Our records indicate that McKesson has received the Atrium Ocean Wet Suction Water Seal Chest Drains from the lot listed on Page 1. " Please examine your inventory and confirm that you have no remaining Atrium Ocean Wet Suction Water Seal Chest Drain 2002-000 with the REF/LOT number listed in this notice. Please return any product that may be affected under this Field Action. " Please respond to this email using the response form on page 3 to acknowledge that you have received this notification. If you have any questions, please contact your Atrium/Getinge representative or call the Atrium/Getinge Customer Support at (603) 880-1433 (press option 1), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone).

Distribution

US Nationwide distribution in the state of VA.

Quantity

24 units (4cases/6)