Remisol Advance Software (Driver Architect) Catalog Number B32129 The Remisol Advance Data Manager (Stand alone Data Management Systems) collects and manages data and manages workflows for connected systems. (i.e. Beckman Coulter Instruments, Automations, LIS &). It provides data analysis capabilities such as automatic results validation, delta checking, reflex testing, quality control, results editing and data management (i.e. archiving and restoration of patient results). The Remisol Advance system also offers workstation consolidation to three Beckman Coulter instruments from a single user console.

Recall: Z-0235-2018 · Initiated October 16, 2017

Recall

Recall Number
Z-0235-2018
Event Number
78394
Firm
Normand Informatique Normand Informatique Sci Git; 57 Grand place; Cs 10451 Arras CEDEX France
FEI Number
3006543086
Product Code
JQP
Status
Terminated
Root Cause
Software design
Initiated
October 16, 2017
Terminated
February 13, 2024

Description

Remisol Advance Software (Driver Architect) Catalog Number B32129 The Remisol Advance Data Manager (Stand alone Data Management Systems) collects and manages data and manages workflows for connected systems. (i.e. Beckman Coulter Instruments, Automations, LIS &). It provides data analysis capabilities such as automatic results validation, delta checking, reflex testing, quality control, results editing and data management (i.e. archiving and restoration of patient results). The Remisol Advance system also offers workstation consolidation to three Beckman Coulter instruments from a single user console.

Reason

Beckman Coulter has identified that due to a software nonconformity in connection with a changed service setting Remisol could display wrong results.

Action

Beckman Coulter sent an Urgent Medical Device Recall letter dated October 16, 2017, to all affected customers. Beckman Coulter has identified a potential issue due to a software nonconformity which could occur after a default service setting (NoCondition) is manually changed from off to on, potentially causing results outside of analytical range to be erroneously displayed. For further questions, please call (800) 526-3821.

Distribution

Worldwide Distribution - US Distribution and to the countries of : Australia, France, Germany, Israel and Russia

Quantity

34 units