Oral Probe (Product code 02893-000), accessory used with Welch Allyn SureTemp Plus 690 (Product codes 01690-200 and 01690-400) Thermometers and Welch Allyn SureTemp Plus 692 (01692-200) thermometers
Recall
- Recall Number
- Z-0233-2026
- Event Number
- 97614
- Firm
- Baxter Healthcare Corporation
- FEI Number
- 1417572
- Product Code
- FLL
- Status
- Open, Classified
- Root Cause
- Process control
- Initiated
- September 17, 2025
- Posted
- October 23, 2025
- Address
- 1 Baxter Pkwy, Deerfield, IL, 60015-4625
Description
Oral Probe (Product code 02893-000), accessory used with Welch Allyn SureTemp Plus 690 (Product codes 01690-200 and 01690-400) Thermometers and Welch Allyn SureTemp Plus 692 (01692-200) thermometers
Affected oral/axillary probes were inadvertently programmed with the rectal probe configuration. Affected probes will potentially display a lower temperature than the true value, which may lead to a delay in managing fever-related symptoms or infection.
Beginning September 17, 2025, firm notified impacted distributors via Medical Device Recall letters. Customers were instructed to return any affected product and submit a request for replacements. Distributors were instructed to notify their end users via a provided Important Product Information Letter. For general questions regarding this action, please contact Welch Allyn technical service at 800-535-6663. Select the following sequence of options: 2 (technical support), 1(English), 1 (physical exam devices), 3 (thermometry), between the hours of 8:00 AM and 8:00 PM Eastern Time, Monday through Friday.
Distribution US nationwide, Canada and Colombia.