FDA Recall Open, Classified

t:slim X2 insulin pump with Interoperable Technology, Control-IQ+ technology, REF: 1014070, 1014071, 1014072, 1014073, and User Guide when using Dexcom G7 sensor; Tandem Mobi insulin pump with Interoperable Technology, Control-IQ+ technology, REF: 1014801, 1014811, 1014813, and User Guide when using Dexcom G7 sensor

Recall: Z-0232-2026 · Initiated August 5, 2025

Recall

Recall Number
Z-0232-2026
Event Number
97403
Firm
Tandem Diabetes Care, Inc.
FEI Number
3007981285
Product Code
QFG
Status
Open, Classified
Root Cause
Software change control
Initiated
August 5, 2025
Posted
October 23, 2025
Address
12400 High Bluff Dr, San Diego, CA, 92130-3077

Description

t:slim X2 insulin pump with Interoperable Technology, Control-IQ+ technology, REF: 1014070, 1014071, 1014072, 1014073, and User Guide when using Dexcom G7 sensor; Tandem Mobi insulin pump with Interoperable Technology, Control-IQ+ technology, REF: 1014801, 1014811, 1014813, and User Guide when using Dexcom G7 sensor

Reason

A software defect in versions 7.9.0.1 and 7.10.1 of the pump software for Tandem t:slim X2 and version 7.9.0.1 of the Tandem Mobi pump with Control-IQ+ technology that are paired with a G7 sensor may result in an unexpected automatic insulin correction bolus (autobolus ) which could result in hypoglycemia.

Action

On 8/5/2025, Field Safety Notices were sent to customers informing them of the following: Actions to avoid the issue: - Wait until your G7 sensor session ends before inserting a new G7 sensor. - However, if you do insert a new G7 sensor before your last sensor session ends, wait to pair until at least 20 minutes after the last session ends. Firm is working on a software update to resolve this issue. If you have questions or need help, please call firm's Customer Support 1-877-801-6901 or email [email protected]. Our team is available 24/7/365. On 9/16/2025, correction notices were emailed to customers who were informed the firm updated user guides to align with instructions in the correction notice. Also, the firm will notify impacted users via an updated Field Correction Notice when the software update is available. The notice will provide instructions on how to remotely update their software.

Distribution

US nationwide distribution including Puerto Rico.

Quantity

122,958