FDA Recall Open, Classified

Olympus Endoscopic Flushing Pump, Model/Catalog Number: OFP-2 (K10001141)

Recall: Z-0229-2025 · Initiated August 23, 2024

Recall

Recall Number
Z-0229-2025
Event Number
95440
Firm
Olympus Corporation of the Americas
FEI Number
2429304
Product Code
FEQ
Status
Open, Classified
Root Cause
Equipment maintenance
Initiated
August 23, 2024
Posted
October 25, 2024
Address
3500 Corporate Pkwy, Center Valley, PA, 18034-8229

Description

Olympus Endoscopic Flushing Pump, Model/Catalog Number: OFP-2 (K10001141)

Reason

During product testing, Olympus identified an intermittent loss of function of the OFP-2 Flushing Pump caused by an internal component connection failure.

Action

An URGENT: MEDICAL DEVICE CORRECTION notification letter dated 9/20/24 was sent to customers. Actions Required: Our records indicate that your facility has purchased one or more of the affected products. The range of affected serial numbers is included at the top of this letter. The serial number is on a label affixed to the back of the OFP-2 pump. An Olympus representative will create a Return Material Authorization for the return of your device to Olympus for an inspection. You may continue to use your OFP-2 pump until an Olympus representative performs the inspection. If the OFP-2 fails the inspection, the OFP-2 will require repair. Additionally, Olympus requires you to take the following actions: 1. Carefully read the content of this notification. 2. Olympus requests that you acknowledge receipt of this letter. Access the Olympus recall portal to indicate that you have received this notification. a. Go to https://olympusamerica.com/recall b. Enter the recall number "0455 c. Complete the form as instructed. 3. If you have further distributed this product, identify your customers, and forward them this notification. 4. Beginning November 2024, please contact Olympus Customer Service at 1-800-848-9024, option 3, to request an Air Switch Assembly Inspection . Olympus will arrange for the return of your device to our Olympus Repair Center at a mutually convenient time. The Olympus representative s record of repair will also serve as the completion record of this action for your facility. Olympus fully appreciates your prompt cooperation in addressing this situation. If you require additional information, please do not hesitate to contact me directly by phone at (647) 999-3203 or by e-mail at [email protected].

Distribution

US Nationwide distribution.

Quantity

445 units