MATResponder Tourniquet; Catalog number MATR: UPC 78353 00008; MATCombat Tourniquet, Catalog number MATC; UPC 78353 00010; MATResponder Tourniquet and MATCombat Tourniquet are essentially the same product with different color schemes manual nonpneumatic tourniquet devices. The device provides for emergency blood flow occlusion to the limbs
Recall
- Recall Number
- Z-0228-2016
- Event Number
- 72096
- Firm
- Pyng Medical Corporation 210-13480 Crestwood Pl Richmond Canada
- FEI Number
- 3003437923
- Product Code
- GAX
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- August 31, 2015
- Terminated
- July 5, 2016
Description
MATResponder Tourniquet; Catalog number MATR: UPC 78353 00008; MATCombat Tourniquet, Catalog number MATC; UPC 78353 00010; MATResponder Tourniquet and MATCombat Tourniquet are essentially the same product with different color schemes manual nonpneumatic tourniquet devices. The device provides for emergency blood flow occlusion to the limbs
Once tested, it is required that the tourniquet be returned to its Ready position then placed in its packaging. Each tourniquet was tested for functionality, but in some cases, the tourniquet was not returned to its Ready position prior to packaging.
PYNG Medical sent Recall Notification letters dated August 28, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed: Please immediately cease use of the MATResponder and MATCombat tourniquet lots specified above. Inform all of your teams to cease use of or distribution of these lots. Please immediately place all MATResponder Tourniquets and MATCombat Tourniquets that you have in inventory into a quarantined location. For further questions please call (604) 303-7964. .
Worldwide Distribution - US Distribution and to the countries of : Colombia, France, Canada, Italy, Germany, Finland, United Kingdom, Australia, The Netherlands, Japan, Hungary, Indonesia, Israel, Mexico, Poland and Spain.