Bag Tee Assembly Model P1407A . An accessory to the dental flowmeter which has a connection function for complete delivery of mixed gases to the breathing circuit portion of the gas scavenging apparatus. The Bag Tee assembly features a non-breathing valve and an Emergency Air Intake located on the bag tee which comply with the American Dental Association guidelines.
Recall
- Recall Number
- Z-0228-2009
- Event Number
- 49492
- Firm
- Porter Instrument Co Inc
- FEI Number
- 2518157
- Product Code
- CAX
- Status
- Terminated
- Root Cause
- Employee error
- Initiated
- August 26, 2008
- Posted
- October 31, 2008
- Terminated
- September 24, 2010
- Address
- 245 Township Line Rd, Hatfield, PA, 19440
Description
Bag Tee Assembly Model P1407A . An accessory to the dental flowmeter which has a connection function for complete delivery of mixed gases to the breathing circuit portion of the gas scavenging apparatus. The Bag Tee assembly features a non-breathing valve and an Emergency Air Intake located on the bag tee which comply with the American Dental Association guidelines.
inverted check value on the bag tee assembly which can cause leaking of mixed N2O/O2 gas
On 9/2/08, the firm issued an Important Field Correction Notice to their customers informing them of the problem and how to determine if their product was defective and needed to be returned. Contact Porter at 1-800-457-2001 and ask for Dental Customer and refer to the Bag Tee Field Correction Notice.
Nationwide
220 units