FDA Recall Open, Classified

Mobilett Mira wireless (VA20) mobile x-ray system(s)

Recall: Z-0223-2024 · Initiated October 5, 2023

Recall

Recall Number
Z-0223-2024
Event Number
93304
Firm
Siemens Medical Solutions USA, Inc
FEI Number
2240869
Product Code
IZL
Status
Open, Classified
Root Cause
Nonconforming Material/Component
Initiated
October 5, 2023
Posted
November 14, 2023
Address
40 Liberty Blvd, Malvern, PA, 19355-1418

Description

Mobilett Mira wireless (VA20) mobile x-ray system(s)

Reason

Due to the incorrect mounting of the buzzer component, there is no acoustic signal during x ray activation. Otherwise, the system is fully functional, and x ray is applied correctly. By means of the additional visual indicator (exposure LED light) the operator can monitor the x ray release. Also, after a successful x ray exposure, the clinical image is displayed on the monitor.

Action

Customer notification by means of a Customer Safety Advisory Notice (via electronic mail or FedEx) alerting customers to the lack of acoustic signal issue and providing instructions to monitor the visual indicator (exposure LED light) to avoid additional, unnecessary x-ray dose. The CSAN also includes the plan for the affected D810 boards. The boards (containing the faulty mounting of the buzzer) will be replaced, free of charge, at the customer site with field corrective action XP042/23/P. There is no software update.

Distribution

US Nationwide - Worldwide

Quantity

4 distributed in U.S.