FDA Recall Terminated

Stryker Transport wheeled stretcher, Model 735 (ST104), Stryker Medical, Portage, MI 49002

Recall: Z-0223-2008 · Initiated November 15, 2004

Recall

Recall Number
Z-0223-2008
Event Number
37214
Firm
Stryker Medical Div. of Stryker Corporation
FEI Number
1831750
Product Code
FPO
Status
Terminated
Root Cause
Component change control
Initiated
November 15, 2004
Posted
November 10, 2007
Terminated
October 27, 2010
Address
3800 E. Centre Ave., Portage, MI, 49002

Description

Stryker Transport wheeled stretcher, Model 735 (ST104), Stryker Medical, Portage, MI 49002

Reason

The brake cams may crack or develop excessive wear, resulting in the brakes not locking the wheels, or the stretcher not being locked firmly in position.

Action

The firm issued an Urgent Medical Device Correction Letter dated 9/24/07, to each consignee informing them of the problems, that the brakes on each affected stretcher needed to be inspected/tested and removed from service if brake problems were discovered, and that a Stryker service representative would be in contact to replace the brakes.

Distribution

Worldwide-USA, Argentina, Australia, Brazil, Canada, Chile, China, Dubai, France, India, Italy, Japan, Germany, Greece, Hong Kong, Korea, Latin America, Malaysia, Mexico, Netherlands, Poland, Portugal, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, and United Kingdom.

Quantity

81,367 (61,708 U.S. and 19,569 ex-U.S.) for all models.