FDA Recall Open, Classified

CADD-Solis Li-ion Rechargeable Battery Packs, REF 21-2160-XX, which are also sold as a component of the following CADD Solis Ambulatory Infusion Systems: 21-2101-0100-50, 21-2101-0200-02, 21-2101-0200-03, 21-2101-0200-06, 21-2101-0200-07, 21-2101-0200-12, 21-2101-0200-14, 21-2101-0200-17, 21-2101-0200-238, 21-2101-249, 21-2101-51, 21-2102-0200-02, 21-2102-249, 21-2102-51, 21-2111-0100-00, 21-2111-0100-50, 21-2111-0100-51, 21-2111-0200-02, 21-2111-0200-03, 21-2111-0200-06, 21-2111-0200-07, 21-2111-0200-08, 21-2111-0200-12, 21-2111-0200-14, 21-2111-0200-17, 21-2111-0200-232, 21-2111-0200-50, 21-2111-0300-00, 21-2111-0300-01, 21-2111-0300-02, 21-2111-0300-03, 21-2111-0300-06, 21-2111-0300-07, 21-2111-0300-08, 21-2111-0300-09, 21-2111-0300-12, 21-2111-0300-14, 21-2111-0300-17, 21-2111-0300-231, 21-2111-0300-232, 21-2111-0300-50, 21-2111-0400-01, 21-2111-0400-51, 21-2111-0401-02L, 21-2111-0401-03L, 21-2111-0401-06L, 21-2111-0401-07, 21-2111-0401-07L, 21-2111-0401-08L, 21-2111-0401-12L, 21-2111-0401-14L, 21-2111-0401-17L, 21-2111-0401-249, 21-2111-0401-51, 21-2111-0401-51L, 21-2111-0401-78, 21-2111-0402-02, 21-2111-0402-02L, 21-2111-0402-03L, 21-2111-0402-06L, 21-2111-0402-07L, 21-2111-0402-08, 21-2111-0402-08L, 21-2111-0402-09_LOAN, 21-2111-0402-09-FUK, 21-2111-0402-09JP, 21-2111-0402-09-KYO, 21-2111-0402-09-NAG, 21-2111-0402-09-OKA, 21-2111-0402-09-OSA, 21-2111-0402-09-SAP, 21-2111-0402-09-SEN, 21-2111-0402-09-TOKE, 21-2111-0402-09-TOKW, 21-2111-0402-09-YOK, 21-2111-0402-12L, 21-2111-0402-14L, 21-2111-0402-17L, 21-2111-0402-249, 21-2111-0402-51, 21-2111-0402-51L, 21-2111-0402-78, 21-2112-0100-50, 21-2112-0100-51, 21-2112-0200-02, 21-2112-0200-06, 21-2112-0300-00, 21-2112-0300-01, 21-2112-0300-02, 21-2112-0300-03, 21-2112-0300-06, 21-2112-0300-231, 21-2112-0300-50, 21-2112-0400-01, 21-2112-0400-51, 21-2112-0401-02L, 21-2112-0401-03L, 21-2112-0401-07L, 21-2112-0401-08L, 21-2112-0401-17L, 21-2112-0401-51, 21-2112-0401-51L, 21-2112-0401-78, 21-2112-0402-02, 21-2112-0402-02L, 21-2112-0402-03L, 21-2112-0402-06L, 21-2112-0402-07L, 21-2112-0402-08L, 21-2112-0402-09JP, 21-2112-0402-09-SAI, 21-2112-0402-09-SEN, 21-2112-0402-09-TOKE, 21-2112-0402-17L, 21-2112-0402-249, 21-2112-0402-51, 21-2112-0402-51L, 21-2120-0100-02, 21-2120-0100-03, 21-2120-0100-06, 21-2120-0100-232, 21-2120-0100-233, 21-2120-0100-50, 21-2120-0100-51, 21-2120-0100-95, 21-2120-0102-02, 21-2120-0102-03, 21-2120-0102-06, 21-2120-0102-07, 21-2120-0102-08, 21-2120-0102-12, 21-2120-0102-13, 21-2120-0102-14, 21-2120-0102-15, 21-2120-0102-17, 21-2120-0102-247, 21-2120-0102-249, 21-2120-0102-51, 21-2120-0102-78, 21-2120-0102-92, 21-2120-0102-98, 21-2120-0103-01, 21-2120-0103-02, 21-2120-0103-03, 21-2120-0103-06, 21-2120-0103-07, 21-2120-0103-08, 21-2120-0103-14, 21-2120-0103-15, 21-2120-0103-17, 21-2120-0103-249, 21-2120-0103-51, 21-2120-0103-78, 21-2120-0104-01, 21-2120-0105-00, 21-2120-0105-01, 21-2120-0105-02L, 21-2120-0105-03L, 21-2120-0105-06L, 21-2120-0105-07L, 21-2120-0105-08L, 21-2120-0105-12L, 21-2120-0105-13L, 21-2120-0105-14L, 21-2120-0105-15L, 21-2120-0105-17L, 21-2120-0105-231, 21-2120-0105-50, 21-2120-0105-50L, 21-2125-0104-01, 21-2125-0105-07L, 21-2125-0105-08L, 21-2125-0105-50L, 21-2127-0104-01, 21-2127-0105-01, 21-2127-0105-02L, 21-2127-0105-03L, 21-2127-0105-06L, 21-2127-0105-08L, 21-2127-0105-50L, 22-4095-0203-51, 22-4095-0204-25, 22-4095-0205-25.

Recall: Z-0222-2025 · Initiated July 16, 2024

Recall

Recall Number
Z-0222-2025
Event Number
95287
Firm
Smiths Medical ASD Inc.
FEI Number
3012307300
Product Code
FRN
Status
Open, Classified
Root Cause
Component design/selection
Initiated
July 16, 2024
Posted
October 31, 2024
Address
6000 Nathan Ln N, Minneapolis, MN, 55442-1690

Description

CADD-Solis Li-ion Rechargeable Battery Packs, REF 21-2160-XX, which are also sold as a component of the following CADD Solis Ambulatory Infusion Systems: 21-2101-0100-50, 21-2101-0200-02, 21-2101-0200-03, 21-2101-0200-06, 21-2101-0200-07, 21-2101-0200-12, 21-2101-0200-14, 21-2101-0200-17, 21-2101-0200-238, 21-2101-249, 21-2101-51, 21-2102-0200-02, 21-2102-249, 21-2102-51, 21-2111-0100-00, 21-2111-0100-50, 21-2111-0100-51, 21-2111-0200-02, 21-2111-0200-03, 21-2111-0200-06, 21-2111-0200-07, 21-2111-0200-08, 21-2111-0200-12, 21-2111-0200-14, 21-2111-0200-17, 21-2111-0200-232, 21-2111-0200-50, 21-2111-0300-00, 21-2111-0300-01, 21-2111-0300-02, 21-2111-0300-03, 21-2111-0300-06, 21-2111-0300-07, 21-2111-0300-08, 21-2111-0300-09, 21-2111-0300-12, 21-2111-0300-14, 21-2111-0300-17, 21-2111-0300-231, 21-2111-0300-232, 21-2111-0300-50, 21-2111-0400-01, 21-2111-0400-51, 21-2111-0401-02L, 21-2111-0401-03L, 21-2111-0401-06L, 21-2111-0401-07, 21-2111-0401-07L, 21-2111-0401-08L, 21-2111-0401-12L, 21-2111-0401-14L, 21-2111-0401-17L, 21-2111-0401-249, 21-2111-0401-51, 21-2111-0401-51L, 21-2111-0401-78, 21-2111-0402-02, 21-2111-0402-02L, 21-2111-0402-03L, 21-2111-0402-06L, 21-2111-0402-07L, 21-2111-0402-08, 21-2111-0402-08L, 21-2111-0402-09_LOAN, 21-2111-0402-09-FUK, 21-2111-0402-09JP, 21-2111-0402-09-KYO, 21-2111-0402-09-NAG, 21-2111-0402-09-OKA, 21-2111-0402-09-OSA, 21-2111-0402-09-SAP, 21-2111-0402-09-SEN, 21-2111-0402-09-TOKE, 21-2111-0402-09-TOKW, 21-2111-0402-09-YOK, 21-2111-0402-12L, 21-2111-0402-14L, 21-2111-0402-17L, 21-2111-0402-249, 21-2111-0402-51, 21-2111-0402-51L, 21-2111-0402-78, 21-2112-0100-50, 21-2112-0100-51, 21-2112-0200-02, 21-2112-0200-06, 21-2112-0300-00, 21-2112-0300-01, 21-2112-0300-02, 21-2112-0300-03, 21-2112-0300-06, 21-2112-0300-231, 21-2112-0300-50, 21-2112-0400-01, 21-2112-0400-51, 21-2112-0401-02L, 21-2112-0401-03L, 21-2112-0401-07L, 21-2112-0401-08L, 21-2112-0401-17L, 21-2112-0401-51, 21-2112-0401-51L, 21-2112-0401-78, 21-2112-0402-02, 21-2112-0402-02L, 21-2112-0402-03L, 21-2112-0402-06L, 21-2112-0402-07L, 21-2112-0402-08L, 21-2112-0402-09JP, 21-2112-0402-09-SAI, 21-2112-0402-09-SEN, 21-2112-0402-09-TOKE, 21-2112-0402-17L, 21-2112-0402-249, 21-2112-0402-51, 21-2112-0402-51L, 21-2120-0100-02, 21-2120-0100-03, 21-2120-0100-06, 21-2120-0100-232, 21-2120-0100-233, 21-2120-0100-50, 21-2120-0100-51, 21-2120-0100-95, 21-2120-0102-02, 21-2120-0102-03, 21-2120-0102-06, 21-2120-0102-07, 21-2120-0102-08, 21-2120-0102-12, 21-2120-0102-13, 21-2120-0102-14, 21-2120-0102-15, 21-2120-0102-17, 21-2120-0102-247, 21-2120-0102-249, 21-2120-0102-51, 21-2120-0102-78, 21-2120-0102-92, 21-2120-0102-98, 21-2120-0103-01, 21-2120-0103-02, 21-2120-0103-03, 21-2120-0103-06, 21-2120-0103-07, 21-2120-0103-08, 21-2120-0103-14, 21-2120-0103-15, 21-2120-0103-17, 21-2120-0103-249, 21-2120-0103-51, 21-2120-0103-78, 21-2120-0104-01, 21-2120-0105-00, 21-2120-0105-01, 21-2120-0105-02L, 21-2120-0105-03L, 21-2120-0105-06L, 21-2120-0105-07L, 21-2120-0105-08L, 21-2120-0105-12L, 21-2120-0105-13L, 21-2120-0105-14L, 21-2120-0105-15L, 21-2120-0105-17L, 21-2120-0105-231, 21-2120-0105-50, 21-2120-0105-50L, 21-2125-0104-01, 21-2125-0105-07L, 21-2125-0105-08L, 21-2125-0105-50L, 21-2127-0104-01, 21-2127-0105-01, 21-2127-0105-02L, 21-2127-0105-03L, 21-2127-0105-06L, 21-2127-0105-08L, 21-2127-0105-50L, 22-4095-0203-51, 22-4095-0204-25, 22-4095-0205-25.

Reason

Damage to the battery pack may have caused a short to a capacitor within the battery pack. While the battery encasement is designed to be flame retardant, a short to the capacitor could lead to melting of the battery pack case. If this issue occurs, the battery pack charging circuit may become inoperable.

Action

Smiths Medical issued an URGENT MEIDCAL DEVICE CORRECTION to its consignees on 07/16/2024 via traceable mailing. The notice explained the issue, potential risk and requested the following actions be taken: 1. Examine the external condition of the battery pack and look for evidence of damage to the outer case. As stated in the battery pack Instructions for Use, if the battery pack housing is cracked or otherwise damaged, replace the battery pack. NEVER use a battery pack that appears damaged. A rechargeable battery pack must be replaced with either another CADD-Solis rechargeable battery pack or with 4 AA batteries. 2. Users with damaged battery packs should submit complaints to: [email protected]; 1-866-216-8806 3. Ensure all users or potential users of these products are immediately made aware of this notification. 4. Complete and return response form to [email protected]. 5. DISTRIBUTORS: If you have distributed potentially affected products to your customers, please immediately forward this notice to them. Request that they complete the response form and return it to [email protected]

Distribution

Worldwide Distribution

Quantity

138039 units