FDA Recall Open, Classified

Arial Water Resistant Pendant with Long Wristband, Models: 59362 and 2560-59362; mobile patient alarm

Recall: Z-0222-2024 · Initiated September 25, 2023

Recall

Recall Number
Z-0222-2024
Event Number
93188
Firm
Securitas Healthcare LLC
FEI Number
1929691
Product Code
IQA
Status
Open, Classified
Root Cause
Component change control
Initiated
September 25, 2023
Posted
November 3, 2023
Address
4600 Vine St, Lincoln, NE, 68503-2823

Description

Arial Water Resistant Pendant with Long Wristband, Models: 59362 and 2560-59362; mobile patient alarm

Reason

Premature battery depletion presents a risk that alarms will not be received by the Arial Wireless Emergency Call System.

Action

Securitas Healthcare issued a recall notice to its consignees on 09/25/2023 via USPS, registered mail. The notice explained the issue, potential hazard, and requested the following actions: 1. Immediately replace Pendant battery upon Low Battery Alert 2. Customer should check Pendant battery status in Arial System application at least daily 3. Customer should check for Pendants that are no longer communicating with the Arial System at least daily This notice must be passed on to those who need to be aware within your organization. - Securitas Healthcare will be replacing all Pendants from the Affected Lots with refurbished Pendants at no cost. The notice has also been posted: https://www.securitashealthcare.com/product-advisories

Distribution

Worldwide distribution - US Nationwide and the country of Canada.

Quantity

49 devices