FDA Recall
Terminated
Proteus XR/a Radiographic X-ray System, models 2259973-1, 2259973-2, 2259973-3, 2259973-4. The system consists of a radiographic table, a overhead tube suspension, high frequency generator and power distribution unit, wall stand, dual focal spot x-ray tube and operator''s console.
Recall: Z-0219-05
·
Initiated December 30, 2004
Recall
- Recall Number
- Z-0219-05
- Event Number
- 30761
- Firm
- General Electric Med Systems LLC
- FEI Number
- 2126677
- Product Code
- KPR
- Status
- Terminated
- Root Cause
- Other
- Initiated
- December 30, 2004
- Posted
- February 14, 2006
- Terminated
- February 21, 2006
- Address
- 3000 N Grandview Blvd, Waukesha, WI, 53188-1615
Description
Proteus XR/a Radiographic X-ray System, models 2259973-1, 2259973-2, 2259973-3, 2259973-4. The system consists of a radiographic table, a overhead tube suspension, high frequency generator and power distribution unit, wall stand, dual focal spot x-ray tube and operator''s console.
Reason
Units have a defect in accordance with 21 CFR 1003.2 because an incorrect exposure can occur.
Action
An 'Urgent: Radiation Safety Notice' letter was issued to hospitals with affected systems describing potential hazards, the planned modification and recommended action to limit the hazard until the planned modification is installed. A GE field Service Representative will install the modification.
Distribution
globally
Quantity
2716