Nihon Kohden Electric Stimulator, Model number: MS-210BK (Optional accessories of MEB-2200A -Neuropack Evoked Potential and EMG Measuring System)
Recall
- Recall Number
- Z-0218-2007
- Event Number
- 36394
- Firm
- Nihon Kohden America Inc
- FEI Number
- 2080783
- Product Code
- GWF
- Status
- Terminated
- Root Cause
- Other
- Initiated
- August 23, 2006
- Posted
- November 28, 2006
- Terminated
- May 23, 2008
- Address
- 90 Icon, Foothill Ranch, CA, 92610-3000
Description
Nihon Kohden Electric Stimulator, Model number: MS-210BK (Optional accessories of MEB-2200A -Neuropack Evoked Potential and EMG Measuring System)
When the device is subject to physical shock, such as a drop to the floor, it may output a different value from the preset value. This may result in a superficial burn on the patient if the stimulation is done for a long time.
All customers will be called by phone, then faxed the Product Notification starting 09/01/2006 Customers were instructed to indicate the serial numbers on each of your MS-210BK stimulators and fax the MS-210BK PRODUCT RECALLREPLACEMENT letter back to Nihon Kohden at: 949-580-15550 or email to [email protected]. Upon receipt of this letter, Nihon Kohden''s customer service department will send a replacement unit.
Nationwide.
97 units in the U.S.