FDA Recall Terminated

Nihon Kohden Electric Stimulator, Model number: MS-210BK (Optional accessories of MEB-2200A -Neuropack Evoked Potential and EMG Measuring System)

Recall: Z-0218-2007 · Initiated August 23, 2006

Recall

Recall Number
Z-0218-2007
Event Number
36394
Firm
Nihon Kohden America Inc
FEI Number
2080783
Product Code
GWF
Status
Terminated
Root Cause
Other
Initiated
August 23, 2006
Posted
November 28, 2006
Terminated
May 23, 2008
Address
90 Icon, Foothill Ranch, CA, 92610-3000

Description

Nihon Kohden Electric Stimulator, Model number: MS-210BK (Optional accessories of MEB-2200A -Neuropack Evoked Potential and EMG Measuring System)

Reason

When the device is subject to physical shock, such as a drop to the floor, it may output a different value from the preset value. This may result in a superficial burn on the patient if the stimulation is done for a long time.

Action

All customers will be called by phone, then faxed the Product Notification starting 09/01/2006 Customers were instructed to indicate the serial numbers on each of your MS-210BK stimulators and fax the MS-210BK PRODUCT RECALLREPLACEMENT letter back to Nihon Kohden at: 949-580-15550 or email to [email protected]. Upon receipt of this letter, Nihon Kohden''s customer service department will send a replacement unit.

Distribution

Nationwide.

Quantity

97 units in the U.S.