Eon 100 system included the following product reference numbers and names: Reference no. 77345- Glucose (GPSL) Reference no. 77450- Uric acid (AUVD) Reference no. 77435- Triglycerides (TGML) Reference no. 77295- Cholesterol (CHVD) Reference no. 77301- Cholesterol HDL (HDLL) Reference no. 77310- Cholesterol, LDL (LDLL)
Recall
- Recall Number
- Z-0214-2016
- Event Number
- 72322
- Firm
- ELITech Clinical Systems SAS Zone lndustrielle SEES France
- FEI Number
- 3007662974
- Product Code
- CGA
- Status
- Terminated
- Root Cause
- Under Investigation by firm
- Initiated
- September 22, 2015
- Posted
- November 5, 2015
- Terminated
- March 4, 2016
Description
Eon 100 system included the following product reference numbers and names: Reference no. 77345- Glucose (GPSL) Reference no. 77450- Uric acid (AUVD) Reference no. 77435- Triglycerides (TGML) Reference no. 77295- Cholesterol (CHVD) Reference no. 77301- Cholesterol HDL (HDLL) Reference no. 77310- Cholesterol, LDL (LDLL)
ELITech clinical Systems reagents based on Trinder reaction used on the Eon 100 instrument system are recalled due to potential drug interference. No complaints were received.
The firm sent recall letters or Technical Bulletins to US consignees on 9/22/15 via registered mail. Technical Bulletin #1022 (for EON 100 users) Technical Bulletin #2008 (for EON 300 users) Technical Bulletin #5064 (for ENVOY users) Technical Bulletin #3009 (for Selectra users) ELITechGroup is not requesting return of the affected product. The method sheets and package inserts will be updated with the new interference claims. When available, the revised labeling will be provided to the customer for customer's use until the new IFU is included in the product packaging. Consignees are requested to return the Confirmation of Receipt form back to the firm by FAX to 401-642-9001 or send email to [email protected] or send via US Post: Vital Diagnostics, 27 Wellington Rd., Lincoln, Rl. If you have any questions, please contact Vital Diagnostics Technical Support at 1-800-345-2822 or 855-ELITECH
US Nationwide Distribution
1389 systems