FDA Recall Terminated

Thoratec Implantable Ventricular Assist Device (IVAD pump) (Component of Thoratec Ventricular Assist Device (VAD) System), Catalog No. 10012-2555-001 ,Thoratec Corporation, Pleasanton, CA. 94588

Recall: Z-0212-2008 · Initiated October 19, 2007

Recall

Recall Number
Z-0212-2008
Event Number
45486
Firm
Thoratec Corp
FEI Number
2916596
Product Code
DSQ
Status
Terminated
Root Cause
Device Design
Initiated
October 19, 2007
Posted
November 21, 2007
Terminated
May 26, 2009
Address
6035 Stoneridge Drive, Pleasanton, CA, 94588

Description

Thoratec Implantable Ventricular Assist Device (IVAD pump) (Component of Thoratec Ventricular Assist Device (VAD) System), Catalog No. 10012-2555-001 ,Thoratec Corporation, Pleasanton, CA. 94588

Reason

Pneumatic Leaks: air leaks developing in the pneumatic driveline when subjected to excessive stress at the junction of the percutaneous line to the pump housing due to sharp bending when implanted (paracorporeal-external-position).

Action

An initial communication was sent to all affected IVAD customers on October 19, 2007, in the form of an "Urgent: Medical Device Correction" notice informing them of the problem. The correction notice requests that its customers/physicians take the following actions: 1) Contact all ongoing patients with IVADs placed in the paracorporeal position to evaluate the percutaneous line adjacent to the pump housing to ensure that it is not subject to extreme flexing or bending at acute angles. 2) Remind implanting surgeons and support staff of proper percutaneous line management and provide the patient with a copy of the recall information provided. Complete and sign an Acknowledgement form and fax it to Thoratec Regulatory Affairs. This is a User Level Recall. The firm also issued a November 7, 2007, Press Statement- Urgent Thoratec Corporation Worldwide Recall of Implantable Ventricular Assist Device.

Distribution

Worldwide: USA including states of AL,AR,AZ,CA,CO, FL,GA, IA,IL, IN,KS,KY,LA, MA, MD, ME,MI, MN, MO,NC, NJ, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, and WI and to the following countries: Austria, Canada, France, Netherlands, Singapore, Switzerland, and The United Kingdom

Quantity

766