Lift Chairs. Electric Positioning Chair.
Recall
- Recall Number
- Z-0212-05
- Event Number
- 29344
- Firm
- Pride Mobility Products Corp
- Product Code
- INO
- Status
- Terminated
- Root Cause
- Other
- Initiated
- June 7, 2004
- Posted
- November 19, 2004
- Terminated
- August 10, 2006
- Address
- 182 Susquehanna Avenue, Exeter, PA, 18643
Description
Lift Chairs. Electric Positioning Chair.
The heating pad on these Lift Chairs may tear at the entry point causing a short and/or overheating, which can damage the chair.
The recalling firm sent a Safety Alert notification to their direct accounts on 5/14/04 informing them of the problem. The direct accounts were instructed to locate the end user and inform them of the problem. Once the chair has been located the direct account is to contact the recalling firm for a replacement and return the recalled chair.
The recalled product was distributed through dealers nationwide. The affected states are as follows: AL, AR, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, LA, OH, OK, MA, MD, MI, MN, MO,MS, NC, ND, NE, NH, NJ, NM, NV, NY, PA, SC, TN, TX, UT, VA, VT, WA, and WV . There is one government account in FL.
532 units