FDA Recall
Terminated
Beckman Coulter Lipase, REF OSR6130 and OSR6230.
Recall: Z-0210-2020
·
Initiated August 24, 2018
Recall
- Recall Number
- Z-0210-2020
- Event Number
- 83583
- Firm
- Beckman Coulter Inc.
- FEI Number
- 2050012
- Product Code
- CDT
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- August 24, 2018
- Terminated
- August 2, 2024
- Address
- 250 S Kraemer Blvd, Brea, CA, 92821-6232
Description
Beckman Coulter Lipase, REF OSR6130 and OSR6230.
Reason
NAPQI, a metabolite of Acetaminophen, causes negative interference with the recalled product assays at levels present in the serum/plasma of overdose patients.
Action
The recalling firm issued letters dated 8/21/2018 via email or postal mail if an email address was not available on 8/24/2018 to their customers. The letter identified the affected product, problem and actions to be taken. For questions call 1-800-223-0139 in the United States and Canada. Outside the United States and Canada contact local Beckman Coulter Representative.
Distribution
Distribution was nationwide. There was government and military distribution. Foreign distribution was made to Canada.
Quantity
20,926 units