FDA Recall Terminated

AMSCO 400 and AMSCO C Small Steam Sterilizers

Recall: Z-0210-2016 · Initiated September 3, 2015

Recall

Recall Number
Z-0210-2016
Event Number
72217
Firm
Steris Corporation
FEI Number
1527821
Product Code
FLE
Status
Terminated
Root Cause
Software design
Initiated
September 3, 2015
Posted
November 4, 2015
Terminated
January 22, 2016
Address
5960 Heisley Rd, Mentor, OH, 44060-1834

Description

AMSCO 400 and AMSCO C Small Steam Sterilizers

Reason

In AMSCO Small Steam Sterilizers equipped with the CS-iQ feature, the sterilizer software inadvertently prevents the capability to operate the foot pedal, which allows the Operator to open and close the sterilizer door.

Action

On 9/21/2015 the firm sent Recall Correction Notices to their customers.

Distribution

US nationwide distribution.

Quantity

32 units