FDA Recall Terminated

Exacta-Mix 2400 pharmacy compounding system.

Recall: Z-0210-04 · Initiated November 7, 2003

Recall

Recall Number
Z-0210-04
Event Number
27695
Firm
Baxa Corporation
FEI Number
1419106
Product Code
NEP
Status
Terminated
Root Cause
Other
Initiated
November 7, 2003
Posted
July 20, 2004
Terminated
February 24, 2004
Address
14445 Grasslands Dr, Englewood, CO, 80112-7062

Description

Exacta-Mix 2400 pharmacy compounding system.

Reason

If device stops due to an alarm or operater pause, inlet valves will remain open, allowing some unwanted mixing of ingredients in the valve chamber..

Action

All users were contacted by phone and fax on 11/7/03 with instructions to safely use the device.

Distribution

Nationwide. Foreign distribution to UK. No government or military consignees.

Quantity

231 systems