FDA Recall Terminated

HENRY SCHEIN Safety Blood Collection Needle 21Gx 1 1/4 Inch, REF 570-2067, 100pcs/box 1000pcs/ctn 350*210*350mm

Recall: Z-0208-2022 · Initiated August 24, 2021

Recall

Recall Number
Z-0208-2022
Event Number
88757
Firm
Zhejiang Kindly Medical Devices Co., Ltd. # 758 Binghai Five Rd. Longwan Dist Wenzhou China
FEI Number
3003560965
Product Code
FMI
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
August 24, 2021
Terminated
July 21, 2023

Description

HENRY SCHEIN Safety Blood Collection Needle 21Gx 1 1/4 Inch, REF 570-2067, 100pcs/box 1000pcs/ctn 350*210*350mm

Reason

Potential for difficulty in safety mechanism activation.

Action

On 8/24/2021, Zhejiang Kindly Medical Devices Co., Ltd. issued Urgent Medical Device Safety Alert notices to two customers. On 9/17/2021, issued a second notice to customers dated 8/24/2021. Action to be Taken by User: 1 - Advise the user to lower the angle between safety protective cap and needle tube by slight pressing in the direction of activation, then using a thumb or hard surface to activate the safety protective cap and lock the needle tube in the IFU, then together with corresponding illustration. 2 - Request the user to strictly follow the guidance of the IFU.

Distribution

U.S. Nationwide distribution.