FDA Recall Terminated

ACT Cellular Phone Monitor; a component of the ACT Ambulatory Cardiac Telemetry; an automatically activated system with a high performance EKG, office hookup and auto send/auto detect that requires no patient intervention to either capture or transmit an arrhythmia when it occurs; Manufactured by Card Guard, 2 Pekeris St, Rehovot, 76100, Israel; there are two models used with either the ACT 1 sensor or the ACT III sensor: a) Model Ozone-HTC b) Model BlackJack2 SGH-i617 LifeStar ACT provides critical ECG information by capturing the onset and termination/escape of an arrhythmia to assist in correctly identifying and treating the patient. Patients are monitored for up to 30 days and all of the data is available at the request of the physician during the service period and for 7 days following the end of service.

Recall: Z-0208-2012 · Initiated September 26, 2011

Recall

Recall Number
Z-0208-2012
Event Number
59990
Firm
Lifewatch Services Inc
FEI Number
1000127138
Product Code
DXH
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
September 26, 2011
Posted
November 17, 2011
Terminated
May 15, 2012
Address
10255 W Higgins Rd, O'Hare International Center II Rosemont, IL, 60018-5606

Description

ACT Cellular Phone Monitor; a component of the ACT Ambulatory Cardiac Telemetry; an automatically activated system with a high performance EKG, office hookup and auto send/auto detect that requires no patient intervention to either capture or transmit an arrhythmia when it occurs; Manufactured by Card Guard, 2 Pekeris St, Rehovot, 76100, Israel; there are two models used with either the ACT 1 sensor or the ACT III sensor: a) Model Ozone-HTC b) Model BlackJack2 SGH-i617 LifeStar ACT provides critical ECG information by capturing the onset and termination/escape of an arrhythmia to assist in correctly identifying and treating the patient. Patients are monitored for up to 30 days and all of the data is available at the request of the physician during the service period and for 7 days following the end of service.

Reason

LifeWatch received complaints of minor shocks and burns during use of the LifeStar ACT systems with shorts in the leads.

Action

The firm, LifeWatch, sent an " Urgent Device Recall" letter dated September 27, 2011 to the affected customers. The letter describes the product, problem and action to be taken. The customers were instructed to immediately refrain from providing these devices to their customers; stop using the affected sensors, return their entire kit in the pre-paid mailer, and use the new device kit included with the recall letter. In addition, the letter informed the users that the replacement device has upgraded software and firmware that allows for detection of shorts in the cable, and if detected, cuts off current flow to the patient to prevent minor shocks and burns. If you have any questions, please call (847) 813-4258.

Distribution

Nationwide distribution.

Quantity

532 phone monitors