FDA Recall
Terminated
AD5 table, modified, used with the XPER x-ray systems. The modification on the table is the cables used for connecting to the XPER systems.
Recall: Z-0208-06
·
Initiated October 12, 2005
Recall
- Recall Number
- Z-0208-06
- Event Number
- 33873
- Firm
- Philips Medical Systems North America Co. Phillips
- FEI Number
- 1000524572
- Product Code
- IZI
- Status
- Terminated
- Root Cause
- Other
- Initiated
- October 12, 2005
- Posted
- November 29, 2005
- Terminated
- March 9, 2011
- Address
- 22100 Bothell Everett Hwy, Bothell, WA, 98021-8431
Description
AD5 table, modified, used with the XPER x-ray systems. The modification on the table is the cables used for connecting to the XPER systems.
Reason
Potential for high voltage exposure involving the injector interface connector of patient support tables.
Action
On 10/10/05 the firm issued a letter to customers alerting customers to the problem and advising that the firm will perform an upgrade to the devices at customer sites. In March 2006 the firm issued a second notification letter involving more units.
Distribution
Devices were distributed nationwide to hospitals/medical centers.
Quantity
see total above