FDA Recall
Terminated
AD6 table for use with Integris and Integris Allura x-ray systems. The AD6 is the tilt version.
Recall: Z-0207-06
·
Initiated October 12, 2005
Recall
- Recall Number
- Z-0207-06
- Event Number
- 33873
- Firm
- Philips Medical Systems North America Co. Phillips
- FEI Number
- 1000524572
- Product Code
- IZI
- Status
- Terminated
- Root Cause
- Other
- Initiated
- October 12, 2005
- Posted
- November 29, 2005
- Terminated
- March 9, 2011
- Address
- 22100 Bothell Everett Hwy, Bothell, WA, 98021-8431
Description
AD6 table for use with Integris and Integris Allura x-ray systems. The AD6 is the tilt version.
Reason
Potential for high voltage exposure involving the injector interface connector of patient support tables.
Action
On 10/10/05 the firm issued a letter to customers alerting customers to the problem and advising that the firm will perform an upgrade to the devices at customer sites. In March 2006 the firm issued a second notification letter involving more units.
Distribution
Devices were distributed nationwide to hospitals/medical centers.
Quantity
529 units with MD6 tablets