FDA Recall Terminated

AD6 table for use with Integris and Integris Allura x-ray systems. The AD6 is the tilt version.

Recall: Z-0207-06 · Initiated October 12, 2005

Recall

Recall Number
Z-0207-06
Event Number
33873
Firm
Philips Medical Systems North America Co. Phillips
FEI Number
1000524572
Product Code
IZI
Status
Terminated
Root Cause
Other
Initiated
October 12, 2005
Posted
November 29, 2005
Terminated
March 9, 2011
Address
22100 Bothell Everett Hwy, Bothell, WA, 98021-8431

Description

AD6 table for use with Integris and Integris Allura x-ray systems. The AD6 is the tilt version.

Reason

Potential for high voltage exposure involving the injector interface connector of patient support tables.

Action

On 10/10/05 the firm issued a letter to customers alerting customers to the problem and advising that the firm will perform an upgrade to the devices at customer sites. In March 2006 the firm issued a second notification letter involving more units.

Distribution

Devices were distributed nationwide to hospitals/medical centers.

Quantity

529 units with MD6 tablets