SCOPIX LR 5200P Laser Imagers, Type 8394/600, serial numbers between 4000 and 5266.

Recall: Z-0207-04 · Initiated November 25, 2003

Recall

Recall Number
Z-0207-04
Event Number
27844
Firm
AGFA Corp.
FEI Number
3001236302
Product Code
LMC
Status
Terminated
Root Cause
Other
Initiated
November 25, 2003
Posted
December 11, 2003
Terminated
October 1, 2004
Address
10 S Academy St, Greenville, SC, 29601-2632

Description

SCOPIX LR 5200P Laser Imagers, Type 8394/600, serial numbers between 4000 and 5266.

Reason

The GS1 circuit board may fail resulting in smoke or fire.

Action

Consignees were notified by letter on 11/25/2003. They were informed of the problem and were advised, if they have the affected type and serial number, to contact the Agfa service representative to arrange for repair. The firm intends to install a modified circuit board on each machine remaining in operation at customer locations. Consignees were instructed to discontinue use of the device immediately and to disconnect the unit from the power supply if any odor of smoke or fire is detected. They were urged to complete the Customer Acknowledgement form included with the letter and return it in the enclosed postage paid envelope.

Distribution

The units were distributed to hospitals and other health facilities located nationwide

Quantity

The units were distributed to hospitals and other health facilities located nationwide