Vector TAS Modular Driver, a manual driver for Vector TAS dental screws.
Recall
- Recall Number
- Z-0205-2013
- Event Number
- 63412
- Firm
- Ormco/Sybronendo
- FEI Number
- 2016150
- Product Code
- DZN
- Status
- Terminated
- Root Cause
- Error in labeling
- Initiated
- December 12, 2011
- Posted
- November 5, 2012
- Terminated
- November 5, 2012
- Address
- 1332 S Lone Hill Ave, Glendora, CA, 91740-5339
Description
Vector TAS Modular Driver, a manual driver for Vector TAS dental screws.
The Instructions For Use (IFUs) for Vector TAS Modular Driver for some foreign language sections incorrectly lists the sterilization temperature as 130C. The correct sterilization temperature should be listed as 132C and/or 270F.
"Urgent - Medical Device Correction" letters dated 12/12/2011 were sent via USPS 1st class mail to all customers and again on 2/21/2012 for recall expansion. The device issue was described and recommended actions were provided. Customers were instructed to acknowledge and complete the Return Form. Questions were referred to (800) 854-1741.
Worldwide distribution, including Nationwide (USA) and the countries of Canada, Austria, Belgium, Bulgaria, Switzerland, Germany, Denmark, France, Estonia, Spain, Finland, Iceland, United Kingdom, Greece, Croatia, Hungary, Ireland, Italy, Luxembourg, Netherlands, Monaco, Poland, Romania, Serbia, Slovenia, Turkey, Jersey, and Martinique.