FDA Recall Terminated

SUPER TORQUE MB Angiographic Catheter product is labeled in part: "***5F (1.65 mm)***110 cm .038" (0.97 mm)***PIG***Max. 1200 psi (8274 kPa)***Markers***Open End 6 Sideholes***Angiographic Catheter*** REF Cat No.***Lot***Use By***Assembled in Mexico Cordis Corporation, 14201 NW 60th Ave. Miami Lakes, Florida 33014, USA***EC***REP*** Cordis Cashel, Cahir Road Cashel, Co Tipperary, Ireland***Cordis***a Johnson&Johnson company***" UPN H739532598A3, Cat No 532-598A, No. MB 10, French Size 5, Shape (PIG), Length cm 110, Flow Rate, ml/sec, 18, Sideholes 6. UPN H739532598B3, Cat No 532-598B, No. MB 20, French Size 5, Shape (PIG), Length cm 110, Flow Rate, ml/sec, 18, Sideholes 6. UPN H73953298C3, Cat No 532-598C, No MB 20 French Size 5, Shape (PIG), Length, cm 65, Flow Rate, ml/sec, 30, Side Holes 6. UPN H73953298D3, Cat No 532-598D, No MB 2 French Size 5, Shape (PIG), Length , cm 70, Flow Rate, ml/sec, 30, Side Holes 10. Cordis SUPER TORQUE MB angiographic catheters with Market Bands are designed to provide angiographic visualization and linear measurement of the vasculature when combined with the delivery of radiopaque contrast media to selected sites in the vascular system.

Recall: Z-0204-2012 · Initiated November 21, 2011

Recall

Recall Number
Z-0204-2012
Event Number
60392
Firm
Cordis Corporation
FEI Number
1016427
Product Code
DQO
Status
Terminated
Root Cause
Error in labeling
Initiated
November 21, 2011
Posted
November 17, 2011
Terminated
July 23, 2013
Address
14201 Nw 60th Ave, Miami Lakes, FL, 33014-2802

Description

SUPER TORQUE MB Angiographic Catheter product is labeled in part: "***5F (1.65 mm)***110 cm .038" (0.97 mm)***PIG***Max. 1200 psi (8274 kPa)***Markers***Open End 6 Sideholes***Angiographic Catheter*** REF Cat No.***Lot***Use By***Assembled in Mexico Cordis Corporation, 14201 NW 60th Ave. Miami Lakes, Florida 33014, USA***EC***REP*** Cordis Cashel, Cahir Road Cashel, Co Tipperary, Ireland***Cordis***a Johnson&Johnson company***" UPN H739532598A3, Cat No 532-598A, No. MB 10, French Size 5, Shape (PIG), Length cm 110, Flow Rate, ml/sec, 18, Sideholes 6. UPN H739532598B3, Cat No 532-598B, No. MB 20, French Size 5, Shape (PIG), Length cm 110, Flow Rate, ml/sec, 18, Sideholes 6. UPN H73953298C3, Cat No 532-598C, No MB 20 French Size 5, Shape (PIG), Length, cm 65, Flow Rate, ml/sec, 30, Side Holes 6. UPN H73953298D3, Cat No 532-598D, No MB 2 French Size 5, Shape (PIG), Length , cm 70, Flow Rate, ml/sec, 30, Side Holes 10. Cordis SUPER TORQUE MB angiographic catheters with Market Bands are designed to provide angiographic visualization and linear measurement of the vasculature when combined with the delivery of radiopaque contrast media to selected sites in the vascular system.

Reason

Cordis Corporation is issuing a Correction for their SUPER TORQUE MB Angiographic Catheter after receiving a report of marker bands dislodged during a procedure. Model Numbers 532598A, 532598B, 532598C, 532598D, SRD5724MB and SRD5727MB are affected by the Cordis Correction.

Action

Cordis Corporation sent a Correction letter dated November 21, 2011 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. Customers are instructed to read the "Description" and "Recommendations" sections, sign and return an enclosed Acknowledgement Form, and to maintain awareness of the communication until the information has been incorporated into the Cordis Super Torque MB Catheter labeling. Contact Cordis at 800-781-0282 for questions regarding this notice.

Distribution

Worldwide Distribution-USA (nationwide) including Puerto Rico.

Quantity

118,690 units