FDA Recall Terminated

PROFlex Transporter, Model numbers 35PST. The catalog number for the model 35PST is PT3565.

Recall: Z-0201-06 · Initiated March 28, 2003

Recall

Recall Number
Z-0201-06
Event Number
34010
Firm
Ferno-Washington Inc
FEI Number
1523574
Product Code
FPO
Status
Terminated
Root Cause
Other
Initiated
March 28, 2003
Posted
November 26, 2005
Terminated
October 29, 2008
Address
70 Weil Way, Wilmington, OH, 45177-9371

Description

PROFlex Transporter, Model numbers 35PST. The catalog number for the model 35PST is PT3565.

Reason

The firm received complaints of the stretchers folding. A fold is when the stretcher lowers on its own to either the next position, or any of the other seven postions of the stretcher.

Action

The firm went out to customers who had complained of a 'fold' and installed the kit on the stretcher that had experienced the fold and on all other PROflexx stretchers owned by the customer. These customers had received a recall letter, dated March 28, 2003. For customers that had not complained of the stretchers 'folding', the firm sent a letter dated June 24, 2003, to all distributors requesting the distributor provide the firm with a list of customers that had received the PROflex stretchers. Once the firm had a list of all customers, they sent them a letter offering them the kit as an enhancement that they could purchase to help reduce the possibility of false locking.

Distribution

The product was distributed nationwide to 27 states: TN, MA, VT, IL, KY, NY, SC, WI, TX, MO, NH, CO, GA, NC, CA, VA, IA, PA, CT, OH, NJ, ID, WV, AL, AR, ME and IN.

Quantity

2188 total stretchers