FDA Recall Terminated

AccessAED Package (without audio record) Model Number: 9100-0150-0

Recall: Z-0200-05 · Initiated November 3, 2004

Recall

Recall Number
Z-0200-05
Event Number
30378
Firm
Access CardioSystems
FEI Number
3003585458
Product Code
MKJ
Status
Terminated
Root Cause
Other
Initiated
November 3, 2004
Posted
November 20, 2004
Terminated
April 12, 2006
Address
150 Baker Avenue, Ste 108, Concord, MA, 01742-2196

Description

AccessAED Package (without audio record) Model Number: 9100-0150-0

Reason

Device may fail to deliver shock due to a faulty circuit board

Action

AccessCardiosystems issued recall letters on 11/3/04 to US Customers by Registered, Certified, Return Receipt requested USPS mail. The foreign mailings made through Federal Express with delivery confirmation.

Distribution

Nationwide Foreign: Spain, Ireland, Chile, Beirut, India, Israel, Singapore

Quantity

2995 units