FDA Recall Terminated

BD PhoenixTM AP, Catalog Number 448010 The BD Phoenix" AP instrument is designed for use with the BD Phoenix" system is intended to standardize ID broth tube inoculum, add AST indicator to the AST broth tube and transfer an aliquot of ID broth to AST broth tubes, as required for preparing samples for use on the BD Phoenix" system, which performs identification and susceptibility testing

Recall: Z-0198-2018 · Initiated June 1, 2017

Recall

Recall Number
Z-0198-2018
Event Number
78102
Firm
Becton Dickinson & Co.
FEI Number
1119779
Product Code
JQX
Status
Terminated
Root Cause
Process control
Initiated
June 1, 2017
Terminated
May 9, 2018
Address
7 Loveton Cir, Sparks, MD, 21152-9212

Description

BD PhoenixTM AP, Catalog Number 448010 The BD Phoenix" AP instrument is designed for use with the BD Phoenix" system is intended to standardize ID broth tube inoculum, add AST indicator to the AST broth tube and transfer an aliquot of ID broth to AST broth tubes, as required for preparing samples for use on the BD Phoenix" system, which performs identification and susceptibility testing

Reason

Potential unexpected movement of robot arm

Action

BD Diagnostic Systems telephoned all customers on June 1, 2017 and sent an Urgent Product Recall letter dated May 31, 2017, and Response Forms to their customers via email. BD recently confirmed a manufacturing issue with the motor that drives movement of the robot arm. The issue may result in the inability of this motor to withstand the lateral forces encountered during normal operations. Should this occur, the instrument will indicate an E20 error-Pipettor step loss during operation and stop functioning. While there is no loss of patient sample, the samples would need to be manually processed until the instrument can be repaired. To limit any disruption in the workload, BD Service will contact customers to schedule the instrument replacement. Customers are advised to complete and return the response form via email to [email protected]. Customers may continue to use the instrument until replacement is performed. Customers with questions may contact BD (Customer/Technical) Support at 1-800-638-8663. For questions regarding this recall call 410-316-4258.

Distribution

Worldwide Distribution - US including CA, NC & TN Internationally: Canada

Quantity

7 units