High voltage generator XTP-8100G and XTPG-8100G for the Infinix-i series interventional angiography system.

Recall: Z-0198-06 · Initiated September 21, 2005

Recall

Recall Number
Z-0198-06
Event Number
33694
Firm
Toshiba American Med Sys Inc
FEI Number
2020563
Product Code
JAA
Status
Terminated
Root Cause
Other
Initiated
September 21, 2005
Posted
November 24, 2005
Terminated
July 12, 2007
Address
2441 Michelle Dr, Tustin, CA, 92781

Description

High voltage generator XTP-8100G and XTPG-8100G for the Infinix-i series interventional angiography system.

Reason

It was found during a study, arcing might occur in the transformer of the high voltage generator, XTP-8100G and XTBP-8100G displaying an error message and interrupting fluoroscopy.

Action

Consignees notification that a local TAMS service representative will schedule the implementation of the replacement transformer.

Distribution

CA, CO, KS, IL, LA, IN, ME, MI, NC, NJ, NY, OH, PA, TX

Quantity

30