High voltage generator XTP-8100G and XTPG-8100G for the Infinix-i series interventional angiography system.
Recall: Z-0198-06
·
Initiated September 21, 2005
Recall
- Recall Number
- Z-0198-06
- Event Number
- 33694
- Firm
- Toshiba American Med Sys Inc
- FEI Number
- 2020563
- Product Code
- JAA
- Status
- Terminated
- Root Cause
- Other
- Initiated
- September 21, 2005
- Posted
- November 24, 2005
- Terminated
- July 12, 2007
- Address
- 2441 Michelle Dr, Tustin, CA, 92781
Description
High voltage generator XTP-8100G and XTPG-8100G for the Infinix-i series interventional angiography system.
Reason
It was found during a study, arcing might occur in the transformer of the high voltage generator, XTP-8100G and XTBP-8100G displaying an error message and interrupting fluoroscopy.
Action
Consignees notification that a local TAMS service representative will schedule the implementation of the replacement transformer.
Distribution
CA, CO, KS, IL, LA, IN, ME, MI, NC, NJ, NY, OH, PA, TX
Quantity
30