FDA Recall Terminated

7" (18 cm) Appx 0.31 ml, Smallbore Pressure Infusion (400psig) Ext Set w/Remv MicroClave Clear, Clamp, NanoClave T-Connector (Purple Ring), Rotating Luer; unit per pouch, 50 multivac pouches per case. UDI:(01)10887709088190(17)241201(30)50(10)4515232;(01)10887709068350(17)250101(30)50(10)4572761 The high pressure extension set is a device used as conduit tubing to deliver fluids from an ~f electronic or manual injector and may also be used as connector tubing to provide a sterile fluid pathway between two devices.

Recall: Z-0197-2021 · Initiated July 6, 2020

Recall

Recall Number
Z-0197-2021
Event Number
86054
Firm
ICU Medical, Inc.
FEI Number
2025816
Product Code
FPA
Status
Terminated
Root Cause
Process control
Initiated
July 6, 2020
Terminated
May 11, 2023
Address
951 Calle, Amanecer San Clemente, CA, 92673-6212

Description

7" (18 cm) Appx 0.31 ml, Smallbore Pressure Infusion (400psig) Ext Set w/Remv MicroClave Clear, Clamp, NanoClave T-Connector (Purple Ring), Rotating Luer; unit per pouch, 50 multivac pouches per case. UDI:(01)10887709088190(17)241201(30)50(10)4515232;(01)10887709068350(17)250101(30)50(10)4572761 The high pressure extension set is a device used as conduit tubing to deliver fluids from an ~f electronic or manual injector and may also be used as connector tubing to provide a sterile fluid pathway between two devices.

Reason

Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.

Action

On July 6, 2020, the firm, icumedical, sent an "URGENT: MEDICAL DEVICE RECALL" notification of the market action to direct and indirect consignees of record via UPS. Stericycle, Inc. was contracted to send letters to consignees. The following is included in the "Required Actions for Users" section of the medical device recall notification: 1) Discontinue the use and distribution of the affected product immediately. Check your inventory and quarantine all affected product at your facility. 2) Inform potential users of the product in your organization of this notification and complete the response form. Return the completed response form to the fax number or e-mail address on the form, even if you do not have the affected product. 3) Return affected product using the return label provided with the letter. Call 1-866-203-5570 (M F, 8am-5pm ET) if you have not received a return label or require additional labels for returning the affected product. The return labels are for single use only. Please do not reproduce. Visit http://expertezlabel.com to request additional labels for returning affected product. To ensure proper and timely credit, follow the instructions on the return label for returning product. Upon receipt of the completed response form and return of the affected product, you will be credited for any product returned. You will only receive credit for product that you return. NOTE: Credits for product purchased through distributor will be credited by the distributor. 4) If you have distributed the product further, immediately notify your accounts that received the product identified in the Affected Product I Table 1 sections of this notification and ask them to call 1-866-203-5570 (M-F, 8am-5pm ET) to obtain a response form. Follow up Actions: Contact Customer Service using the information provided below for assistance reordering replacement product and for further inquiries: Global Complaint Management -1-844-654-7780 or ProductCompla

Distribution

Worldwide Distribution: US (nationwide): AL, AZ, CA, CO, DE, FL,GA,ID, IL,IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NH,NJ,NM,NV, NY, OH, OK, OR, PA,RI, TN, TX, UT, VA, WA, WI, and WV; and OUS countries of: Belgium, Brazil, Canada, France, Germany, Ireland, Japan, Martinique, Mexico, Poland, and Saudi Arabia.