FDA Recall Terminated

SIGMA SPECTRUM Infusion Pump with Master Drug Library. Intended to be used for the controlled administration of intravenous fluids.

Recall: Z-0197-2014 · Initiated June 10, 2013

Recall

Recall Number
Z-0197-2014
Event Number
66120
Firm
Baxter Healthcare Corp.
FEI Number
1417572
Product Code
FRN
Status
Terminated
Root Cause
Process design
Initiated
June 10, 2013
Posted
November 8, 2013
Terminated
April 20, 2017
Address
1 Baxter Pkwy, Deerfield, IL, 60015-4625

Description

SIGMA SPECTRUM Infusion Pump with Master Drug Library. Intended to be used for the controlled administration of intravenous fluids.

Reason

Baxter Healthcare Corporation issued an Urgent Device Correction for the SIGMA SPECTRUM Infusion Pump with Master Drug Drug Library due to reports of upstream occlusion alarms and air-in-line alarms occurring during infusions.

Action

On June 10, 2013 Baxter Healthcare Corporation issued an Urgent Device Correction for the SIGMA SPECTRUM Infusion Pump with Master Drug Library. The notices were dated June 5, 2013 and were distributed via first class mail. Until further information is provided, users should continue to follow the SIGMA SPECTRUM Infusion Pump Operator's Manual, which identifies a number of environmental and clinical usage factors that contribute to upstream occlusion and air-in-line alarms. For high-risk patients, consider having additional devices as backup. If the biomedical engineer finds that the sensor is not functioning properly, contact Baxter Healthcare at 800-356-3454 (choose option 1). For clinical questions, please contact Medical Information Services at Baxter at 800-933-0303, Monday through Friday 8:00am to 5:00 pm Central Time.

Distribution

Worldwide Distribution-USA including Puerto Rico and Canada.

Quantity

273,835 units