SIGMA SPECTRUM Infusion Pump with Master Drug Library. Intended to be used for the controlled administration of intravenous fluids.
Recall
- Recall Number
- Z-0197-2014
- Event Number
- 66120
- Firm
- Baxter Healthcare Corp.
- FEI Number
- 1417572
- Product Code
- FRN
- Status
- Terminated
- Root Cause
- Process design
- Initiated
- June 10, 2013
- Posted
- November 8, 2013
- Terminated
- April 20, 2017
- Address
- 1 Baxter Pkwy, Deerfield, IL, 60015-4625
Description
SIGMA SPECTRUM Infusion Pump with Master Drug Library. Intended to be used for the controlled administration of intravenous fluids.
Baxter Healthcare Corporation issued an Urgent Device Correction for the SIGMA SPECTRUM Infusion Pump with Master Drug Drug Library due to reports of upstream occlusion alarms and air-in-line alarms occurring during infusions.
On June 10, 2013 Baxter Healthcare Corporation issued an Urgent Device Correction for the SIGMA SPECTRUM Infusion Pump with Master Drug Library. The notices were dated June 5, 2013 and were distributed via first class mail. Until further information is provided, users should continue to follow the SIGMA SPECTRUM Infusion Pump Operator's Manual, which identifies a number of environmental and clinical usage factors that contribute to upstream occlusion and air-in-line alarms. For high-risk patients, consider having additional devices as backup. If the biomedical engineer finds that the sensor is not functioning properly, contact Baxter Healthcare at 800-356-3454 (choose option 1). For clinical questions, please contact Medical Information Services at Baxter at 800-933-0303, Monday through Friday 8:00am to 5:00 pm Central Time.
Worldwide Distribution-USA including Puerto Rico and Canada.
273,835 units