FDA Recall Terminated

The 5085 is a general surgical table with high patient weight capacity, extended width capability, and lower minimal table top elevation.

Recall: Z-0196-2015 · Initiated October 10, 2014

Recall

Recall Number
Z-0196-2015
Event Number
69472
Firm
Steris Corporation
FEI Number
1527821
Product Code
GDC
Status
Terminated
Root Cause
Device Design
Initiated
October 10, 2014
Posted
November 7, 2014
Terminated
June 26, 2015
Address
5960 Heisley Rd, Mentor, OH, 44060-1834

Description

The 5085 is a general surgical table with high patient weight capacity, extended width capability, and lower minimal table top elevation.

Reason

The 5085 and 5085SRT stainless steel column shrouds may become damaged as a result of misalignment during extreme table articulation, excessive side force on the table column, or pressure on the base of the table. Severe damage to the shrouds may prevent the user from raising or lowering the surgical table.

Action

STERIS will mail Consignee Notification Letters to the affected Consignees via FedEx with tracking numbers for delivery confirmation. The Consignee Notification mailing will initiate the week of October 13, 2014.

Distribution

US distribution: AL, AZ, AR, CA, CO, CN, DC, FL, GA, HI, ID, IL, IN, IA, KN, KY, LA, ME, MD, MI, MN, MS, MO, NB, NV, NY, NC, ND, OH, OK, OR, PN, RI, SC, SD, TN, TX, VA, WV, WI. Foriegn: Austrialia, Brazil, Canada, China, Ecuador, India, Korea, Mexico, Pnama, and Thailand.

Quantity

999