FDA Recall Terminated

26" Vision Elect HDTV Surgical Viewing Monitor, model number 0240030960, Manufactured by Stryker Endoscopy San Jose Intended to be used as a visualization device for events occurring in the sterile field for endoscopic surgical procedures.

Recall: Z-0196-2009 · Initiated September 22, 2008

Recall

Recall Number
Z-0196-2009
Event Number
49549
Firm
Stryker Endoscopy
FEI Number
2936485
Product Code
GCJ
Status
Terminated
Root Cause
Device Design
Initiated
September 22, 2008
Posted
October 24, 2008
Terminated
December 21, 2010
Address
5900 Optical Ct, San Jose, CA, 95138

Description

26" Vision Elect HDTV Surgical Viewing Monitor, model number 0240030960, Manufactured by Stryker Endoscopy San Jose Intended to be used as a visualization device for events occurring in the sterile field for endoscopic surgical procedures.

Reason

Knob parts stick together, resulting in poor functionality of the knob.

Action

Urgent: Device Correction letters dated 9/22/2008 were sent to Sales representatives and all accounts. Sales reps were also notified via email. Each Sales Rep was provided rework instructions, a list of affected accounts and serial numbers, and new knobs to install on affected units. Customers were asked to inspect their inventory for the part number, they were informed that their sales rep would be responsible for replacing the rotary knobs on all affected units, and that they should contact their sales rep to schedule the knob replacement.

Distribution

Worldwide Distribution --- USA including states of Alabama, Arkansas, Arizona, California, Colorado, Connecticut, DC, Delaware, Florida, Georgia, Hawaii, Iowa, Idaho, Illinois, Indiana, Kansas, Kentucky, Louisiana, Massachusetts, Maryland, Michigan, Minnesota, Missouri, Montana, North Carolina, North Dakota, Nebraska, New Jersey, New Mexico, Nevada, New York, Ohio, Oklahoma, Oregon, Pennsylvania, Rhode Island, South Carolina, South Dakota, Tennessee, Texas, Utah, Virginia, Vermont, Washington, Wisconsin, West Virginia, and Wyoming, and countries of Australia, Canada, China, England, France, Germany, Greece, Iberia, India, Israel, Japan, Korea, Mexico, Netherlands, Poland, Portugal, Scandinavia, Singapore, South Africa, Switzerland, and Taiwan.

Quantity

2265 devices