FDA Recall Terminated

IBO Blade, Right, for use with Stryker TPS Reciprocating Saw, REF 5100-37-901, Stryker Instruments, Kalamazoo, MI. Product Usage: The intended use of these devices is to cut bone and hard tissue in maxillofacial procedures and have been specifically designed for osteotomies of the mandibular ramus.

Recall: Z-0195-2012 · Initiated October 18, 2011

Recall

Recall Number
Z-0195-2012
Event Number
60145
Firm
Stryker Instruments Div. of Stryker Corporation
FEI Number
1811755
Product Code
DZJ
Status
Terminated
Root Cause
Device Design
Initiated
October 18, 2011
Posted
November 16, 2011
Terminated
July 12, 2013
Address
4100 E Milham Ave, Portage, MI, 49002-9704

Description

IBO Blade, Right, for use with Stryker TPS Reciprocating Saw, REF 5100-37-901, Stryker Instruments, Kalamazoo, MI. Product Usage: The intended use of these devices is to cut bone and hard tissue in maxillofacial procedures and have been specifically designed for osteotomies of the mandibular ramus.

Reason

There is a potential for these blades to break where the blade is welded on to the arbor. This hazard can result in the following harms: tissue damage due to the broken blade being aspirated or swallowed, additional surgery to remove a broken blade, infection due to an unretrieved device component, complications associated with increased time under anesthesia to retrieve a broken blade, osteomyeli

Action

The firm, Stryker, sent an "URGENT MEDICAL DEVICE RECALL NOTIFICATION" letter dated October 18, 2011 to Risk Managers, Office Managers, and Oral Surgeons. The letter described the products under recall, the problem and a series of actions to be taken. Customers were asked to immediately locate recalled products and quarantine. Products were to be shipped back to Stryker. The letter included a Business Reply Form to be faxed back to Stryker Instruments Regulatory Department, 866-521-2762. The letter instructs to forward this notification to all affected locations if the recalled products were further distributed. Customers with questions can call Stryker at 1-800-253-3210.

Distribution

Worldwide Distribution - USA (nationwide) and the countries of Brazil, Canada, China, Germany, Greece, Hong Kong, India, Jordan, Miramar, Saudi Arabia, Singapore, Spain, Switzerland, United Kingdom.

Quantity

5,691