FDA Recall Terminated

Axiom Artis Zee. Angiographic X-ray System.

Recall: Z-0194-2009 · Initiated August 22, 2008

Recall

Recall Number
Z-0194-2009
Event Number
49529
Firm
Siemens Medical Solutions USA, Inc
FEI Number
2240869
Product Code
IZI
Status
Terminated
Root Cause
Labeling design
Initiated
August 22, 2008
Posted
November 4, 2008
Terminated
September 23, 2009
Address
51 Valley Stream Pkwy, Malvern, PA, 19355

Description

Axiom Artis Zee. Angiographic X-ray System.

Reason

Patients could possibly fall off the table if not properly secured when: 1) moving them onto the patient table, 2) moving them on the patient table or 3) removing them from the patient table.

Action

The recalling firm issued a Customer Safety Advisory Notice dated 7/04/08 (including an Addendum to the Operators Manual) which alerts the operator of possible injury to patients if they are not properly secured to the table during examinations and describes the recommended fastening procedures required to properly secure the patient to the table. For questions or additional information, contact Siemens Medical Solutions, Inc. at 610-448-4634.

Distribution

Nationwide Distribution.

Quantity

108 units