FDA Recall Open, Classified

EVIS EXERA b BRONCHOVIDEOSCOPE OLYMPUS Model Number BF-XT190

Recall: Z-0193-2024 · Initiated October 12, 2023

Recall

Recall Number
Z-0193-2024
Event Number
93052
Firm
Olympus Corporation of the Americas
FEI Number
2429304
Product Code
EOQ
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
October 12, 2023
Posted
November 9, 2023
Address
3500 Corporate Pkwy, Center Valley, PA, 18034-8229

Description

EVIS EXERA b BRONCHOVIDEOSCOPE OLYMPUS Model Number BF-XT190

Reason

There is a risk of endobronchial combustion if high-frequency cauterization is performed while supplying oxygen [and/or] the electrode section of the electrosurgical accessory is too close to the distal end of the endoscope.

Action

An "URGENT MEDICAL DEVICE CORRECTIVE ACTION" notification letter dated 10/12/23 was sent to customers. Actions to be taken by the end user: Our records indicate that your facility has purchased one or more of the affected bronchoscopes. Olympus requests you to take the following actions: 1. Inspect your facility for the referenced devices and ensure all personnel are completely knowledgeable and thoroughly aware of the Warnings in affected bronchoscope s Operation Manual for use with High-frequency therapy equipment and that Olympus high-frequency compatible bronchoscopes are compatible only with Combination equipment list in operation manual. You may continue to use the device according to the existing instructions and warnings contained in the Instruction for Use. 2. Complete the enclosed response form and return to our recall partner, Sedgwick, via Email ([email protected]) or Fax (866-808-1177). For any questions about the acknowledgement form, please call the Sedgwick team at 855-215-5128. 3. If you have further distributed this product outside of your facility, forward them this notification and appropriately document your notification process. Olympus requests that you report to Olympus complaints, including any injuries associated with procedures involving energy devices used with Olympus bronchoscopes. Please report complaints to the Olympus Technical Assistance Center (TAC) at 1-800-848-9024 (option 1). Adverse events experienced with the use of this product may also be reported to the Food and Drug Administration MedWatch Adverse Event Reporting program either online, by regular mail or by fax. Olympus regrets any inconvenience caused and fully appreciates your cooperation in this matter. Please do not hesitate to contact me directly at (647) 999-3203 or at [email protected] for any additional information or support concerning this matter.

Distribution

US Nationwide distribution

Quantity

719 devices