BIAD Classic 24 (24 inch collimator). The collimator is a component of the Trionix Biad Full Imaging System.
Recall: Z-0193-2007
·
Initiated August 1, 2006
Recall
- Recall Number
- Z-0193-2007
- Event Number
- 36381
- Firm
- Trionix Research Laboratory Inc
- FEI Number
- 1527587
- Product Code
- KPS
- Status
- Terminated
- Root Cause
- Other
- Initiated
- August 1, 2006
- Posted
- November 16, 2006
- Terminated
- October 29, 2008
- Address
- 8037 Bavaria Rd, Twinsburg, OH, 44087-2261
Description
BIAD Classic 24 (24 inch collimator). The collimator is a component of the Trionix Biad Full Imaging System.
Reason
The ball screw on the unit which supports the detector head (which weighs approximately 800 lbs.) could snap resulting in the detector head dropping on the patient.
Action
The recalling firm modified one of the devices in 2002. The remaining device users were telephoned on/about 11/02/2006. To date the recalling firm has not taken any corrective action regarding devices with serial numbers B-102, B-103, B-104 or B-106 thru B-184.
Distribution
Worldwide, including USA, Australia, Egypt, UK, Belgium.
Quantity
76 devices