FDA Recall Terminated

BIAD Classic 24 (24 inch collimator). The collimator is a component of the Trionix Biad Full Imaging System.

Recall: Z-0193-2007 · Initiated August 1, 2006

Recall

Recall Number
Z-0193-2007
Event Number
36381
Firm
Trionix Research Laboratory Inc
FEI Number
1527587
Product Code
KPS
Status
Terminated
Root Cause
Other
Initiated
August 1, 2006
Posted
November 16, 2006
Terminated
October 29, 2008
Address
8037 Bavaria Rd, Twinsburg, OH, 44087-2261

Description

BIAD Classic 24 (24 inch collimator). The collimator is a component of the Trionix Biad Full Imaging System.

Reason

The ball screw on the unit which supports the detector head (which weighs approximately 800 lbs.) could snap resulting in the detector head dropping on the patient.

Action

The recalling firm modified one of the devices in 2002. The remaining device users were telephoned on/about 11/02/2006. To date the recalling firm has not taken any corrective action regarding devices with serial numbers B-102, B-103, B-104 or B-106 thru B-184.

Distribution

Worldwide, including USA, Australia, Egypt, UK, Belgium.

Quantity

76 devices