FDA Recall Terminated

Size 8 Pedicle Screws of the Allez Spine Laguna Pedicle Screw System, Screw Size (diameter x length) 8 x 80, Catalog Number LG-PR0880, Allez Spine, LLC., Irvine, CA

Recall: Z-0191-2008 · Initiated July 30, 2007

Recall

Recall Number
Z-0191-2008
Event Number
44911
Firm
Allez Spine, LLC
FEI Number
3005032381
Product Code
MNI
Status
Terminated
Root Cause
Device Design
Initiated
July 30, 2007
Posted
March 11, 2008
Terminated
April 9, 2012
Address
2301 Dupont Drive, Suite 510, Irvine, CA, 92612

Description

Size 8 Pedicle Screws of the Allez Spine Laguna Pedicle Screw System, Screw Size (diameter x length) 8 x 80, Catalog Number LG-PR0880, Allez Spine, LLC., Irvine, CA

Reason

Screw Shank Failure; separation of the screw shank portion from the pedicle head portion of the Size 8 Pedicle Screw, when torquing down the single piece locking nut during implantation.

Action

On July 30, 2007 Allez Spine notified all its field representatives and distributors via email of a potential for failure of the Size 8 Pedicle Screws when used in combination with the single piece locking nuts and instructed them to return all unused inventory of Size 8 Screws and the two piece Locking Nuts with saddle. This action was to be carried out to the retail level-hospital, clinic, lab, and physicians sold the product. The email also had an attachment of a list of expected frequently asked questions (FAQ) from surgeons and answers to help them respond. A second Notice to all users was sent by the recalling firm on Feb 29, 2008, to request follow-up of implanted units.

Distribution

Nationwide: including the states of AZ, CA, CO, GA, NV, TX and UT

Quantity

14