FDA Recall Open, Classified

Arietta 850 Ultrasound- Software Version: V1.0.0 through V4.1.3

Recall: Z-0190-2022 · Initiated October 14, 2021

Recall

Recall Number
Z-0190-2022
Event Number
88840
Firm
FujiFilm Healthcare Americas Corporation
FEI Number
1528028
Product Code
IYN
Status
Open, Classified
Root Cause
Software design
Initiated
October 14, 2021
Address
1959 Summit Commerce Park, Twinsburg, OH, 44087-2371

Description

Arietta 850 Ultrasound- Software Version: V1.0.0 through V4.1.3

Reason

Diagnostic ultrasound system with the specified software version and used in conjunction with a Fujifilm (formally Hitachi) transducer does not display the measurement results correctly with two cardiology measurement functions: Mitral Regurgitation Flow (MR Flow) Measurement; Mitral Valve (MV) Measurement. In addition there is an error in the instruction manual.

Action

Fujifi Film issued Medical Device Correction Notice letter dated 10/07/21 via courier on 10/14/21. Letter states reason for recall, health risk and action to take: Customers/Users are advised to accept the corrected software and manual as soon as available. Advisory Notification applies to all customers and distributors of the ARIETTA 850 systems with software version V1.0.0 through V4.1.3, See following page for complete list of affected systems. We will contact you to set up time to update the software and provide the revised manual. (7) Other Information: For questions, comments, or concerns, please contact: Aaron Pierce Director, QA/RA 1959 Summit Commerce Park Twinsburg, OH 44087 Phone (330) 425-1313 x2934, Fax (330) 963-0749 Email [email protected]

Distribution

Nationwide

Quantity

38 units