GRIPPER Needles sold as part of PORT-A-CATH II SINGLE LUMEN SYSTEM. Labeled as the following item descriptions: 1. TRAY, PAC, VEN, TI, PU, 1.9MM, UL, 6FR/INTRO 1/EA; 2. TRAY, PAC, VEN LP, TI, PU, 1.9MM, UL, 6 FR/INTRO 1/EA; 3. TRAY, PAC II, VEN, 2.8MM, CS, 9FR/INTRO 1/EA; 4. TRAY, PAC II, VEN, TI, PU, 1.9MM, UL, 6FR/INTRO 1/EA; 5. TRAY, PAC II, VEN, PU, 2.6MM, UL, 8.5FR/INTRO 1/EA; 6. TRAY, PAC II, VEN, PU, 2.6MM, PRE-AY, 8.5FR/INTRO 1/EA; 7. TRAY, PAC II, VEN LP, PU, 2.6MM, UL, 8.5FR/INTRO, VALVED 1/EA; 8. TRAY, PAC II, VEN LP, PU, 1.9MM, UL, 6FR/INTRO, VALVED 1/EA; 9. TRAY, PAC II, VEN LP, PU, 2.6MM, UL, 8.5FR/INTRO 1/EA; 10. TRAY, PAC II, VEN LP, PU, 2.6MM, PRE-AY, 8.5 FR/INTRO 1/EA; 11. TRAY, PAC II, VEN LP, PU, 1.9MM, UL, 6FR/INTRO 1/EA; 12. TRAY, PROPORT, PLASTIC, 2PC, STD, SGL LMN, 1.0, PU 1/EA; 13. TRAY, PROPORT, PLASTIC, 2PC, STD, SGL LMN, 1.6, PU 1/EA; 14. TRAY, PROPORT, PLASTIC, 2PC, LP, SGL LMN, 1.6, PU 1/EA; 15. TRAY, PROPORT, PLASTIC, 2PC, LP, SGL LMN, 1.0, PU 1/EA; 16. TRAY, PROPORT, PLASTIC, 2PC, LP, SGL LMN, 1.0, SIL 1/EA; 17. TRAY, POWER PAC, VEN, TI, PU, 2.6MM, UL, 8.5 FR/INTRO 1/EA; 18. TRAY, POWER PAC, VEN, TI, PU, 1.9MM, UL, 6 FR/INTRO 1/EA; 19. TRAY, POWER PAC, VEN LP, TI, PU, 1.9MM, UL, 6 FR/INTRO 1/EA; 20. TRAY, POWER PAC II, VEN, TI, PU, 1.9MM, UL, 6 FR/INTRO 1/EA; 21. TRAY, POWER PAC II, VEN, PU, 2.6MM, UL 8.5 FR/INTRO 1/EA; 22. TRAY, POWER PAC II, VEN, PU, 2.6MM, PRE-AY, 8.5 FR/INTRO 1/EA; 23. TRAY, POWER PAC II, VEN LP, PU, 2.6MM, UL 8.5 FR/INTRO 1/EA; 24. TRAY, POWER PAC II, VEN, PU, 2.6MM, UL 8.5 FR/INTRO, VALVED 1/EA; 25. TRAY, POWER PAC II, VEN LP, PU, 2.6MM, UL 8.5 FR/INTRO, VALVED 1/EA; 26. TRAY, POWER PAC II, VEN, PU, 1.9MM, UL, 6 FR/INTRO, VALVED 1/EA; 27. TRAY, POWER PAC II, VEN, LP, PU, 1.9MM, UL, 6 FR/INTRO, VALVED 1/EA; 28. TRAY, POWER PAC II, VEN LP, PU, 1.9MM, UL, 6 FR/INTRO 1/EA; 29. TRAY, POWER PAC II, VEN LP, PU, 1.9MM, PRE-AY, 6 FR/INTRO 1/EA; 30. TRAY, POWER PAC, PASPORT T2, PU, 1.9MM, UL, 6 FR/INTRO 1/EA; 31. TRAY, PAC II, DL, LP, PU, 2.2MM, WING-LOCK, 7FR/INTRO 1/EA; 32. TRAY, POWER PAC II, DL, LP, PU, 2.2MM, WING-LOCK, 7FR, INTRO 1/EA; 33. TRAY, POWERPAC II, DL, LP, PU, 3.2MM, WING-LOCK, 10 FR/INTRO 1/EA; 34. TRAY, POWER PAC II, DL, LP, PU, 3.2MM, WING-LOCK, 10 FR/INTRO, VALVED 1/EA;
Recall
- Recall Number
- Z-0190-2020
- Event Number
- 83286
- Firm
- Smiths Medical ASD Inc.
- FEI Number
- 3012307300
- Product Code
- LJT
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- July 8, 2019
- Terminated
- August 23, 2020
- Address
- 6000 Nathan Ln N, Minneapolis, MN, 55442-1690
Description
GRIPPER Needles sold as part of PORT-A-CATH II SINGLE LUMEN SYSTEM. Labeled as the following item descriptions: 1. TRAY, PAC, VEN, TI, PU, 1.9MM, UL, 6FR/INTRO 1/EA; 2. TRAY, PAC, VEN LP, TI, PU, 1.9MM, UL, 6 FR/INTRO 1/EA; 3. TRAY, PAC II, VEN, 2.8MM, CS, 9FR/INTRO 1/EA; 4. TRAY, PAC II, VEN, TI, PU, 1.9MM, UL, 6FR/INTRO 1/EA; 5. TRAY, PAC II, VEN, PU, 2.6MM, UL, 8.5FR/INTRO 1/EA; 6. TRAY, PAC II, VEN, PU, 2.6MM, PRE-AY, 8.5FR/INTRO 1/EA; 7. TRAY, PAC II, VEN LP, PU, 2.6MM, UL, 8.5FR/INTRO, VALVED 1/EA; 8. TRAY, PAC II, VEN LP, PU, 1.9MM, UL, 6FR/INTRO, VALVED 1/EA; 9. TRAY, PAC II, VEN LP, PU, 2.6MM, UL, 8.5FR/INTRO 1/EA; 10. TRAY, PAC II, VEN LP, PU, 2.6MM, PRE-AY, 8.5 FR/INTRO 1/EA; 11. TRAY, PAC II, VEN LP, PU, 1.9MM, UL, 6FR/INTRO 1/EA; 12. TRAY, PROPORT, PLASTIC, 2PC, STD, SGL LMN, 1.0, PU 1/EA; 13. TRAY, PROPORT, PLASTIC, 2PC, STD, SGL LMN, 1.6, PU 1/EA; 14. TRAY, PROPORT, PLASTIC, 2PC, LP, SGL LMN, 1.6, PU 1/EA; 15. TRAY, PROPORT, PLASTIC, 2PC, LP, SGL LMN, 1.0, PU 1/EA; 16. TRAY, PROPORT, PLASTIC, 2PC, LP, SGL LMN, 1.0, SIL 1/EA; 17. TRAY, POWER PAC, VEN, TI, PU, 2.6MM, UL, 8.5 FR/INTRO 1/EA; 18. TRAY, POWER PAC, VEN, TI, PU, 1.9MM, UL, 6 FR/INTRO 1/EA; 19. TRAY, POWER PAC, VEN LP, TI, PU, 1.9MM, UL, 6 FR/INTRO 1/EA; 20. TRAY, POWER PAC II, VEN, TI, PU, 1.9MM, UL, 6 FR/INTRO 1/EA; 21. TRAY, POWER PAC II, VEN, PU, 2.6MM, UL 8.5 FR/INTRO 1/EA; 22. TRAY, POWER PAC II, VEN, PU, 2.6MM, PRE-AY, 8.5 FR/INTRO 1/EA; 23. TRAY, POWER PAC II, VEN LP, PU, 2.6MM, UL 8.5 FR/INTRO 1/EA; 24. TRAY, POWER PAC II, VEN, PU, 2.6MM, UL 8.5 FR/INTRO, VALVED 1/EA; 25. TRAY, POWER PAC II, VEN LP, PU, 2.6MM, UL 8.5 FR/INTRO, VALVED 1/EA; 26. TRAY, POWER PAC II, VEN, PU, 1.9MM, UL, 6 FR/INTRO, VALVED 1/EA; 27. TRAY, POWER PAC II, VEN, LP, PU, 1.9MM, UL, 6 FR/INTRO, VALVED 1/EA; 28. TRAY, POWER PAC II, VEN LP, PU, 1.9MM, UL, 6 FR/INTRO 1/EA; 29. TRAY, POWER PAC II, VEN LP, PU, 1.9MM, PRE-AY, 6 FR/INTRO 1/EA; 30. TRAY, POWER PAC, PASPORT T2, PU, 1.9MM, UL, 6 FR/INTRO 1/EA; 31. TRAY, PAC II, DL, LP, PU, 2.2MM, WING-LOCK, 7FR/INTRO 1/EA; 32. TRAY, POWER PAC II, DL, LP, PU, 2.2MM, WING-LOCK, 7FR, INTRO 1/EA; 33. TRAY, POWERPAC II, DL, LP, PU, 3.2MM, WING-LOCK, 10 FR/INTRO 1/EA; 34. TRAY, POWER PAC II, DL, LP, PU, 3.2MM, WING-LOCK, 10 FR/INTRO, VALVED 1/EA;
Potential for certain GRIPPER Needles to contain an occluded or blocked needle.
Smiths Medical notified customers during the week of 07/08/2019 via "URGENT MEDICAL DEVICE RECALL NOTICE" letter. The Recall Notice advised customers that Smiths Medical has initiated a voluntary field removal for specific lots of GRIPPER needles and specific lots of PORT-A-CATH Trays containing GRIPPER needles. Customers were instructed to examine inventory for affected product and complete and return the Response Form to [email protected]. Once the response form is received, Customer Service will provide a shipping label to return all affected product to Stericycle. A copy of the completed Response Form must be included inside EACH BOX of returned product to facilitate processing. Ensure boxes are sealed and labeled with your facility name prior to shipping. Distributors were instructed to immediately notify customers if the product was further distributed. A copy of the Recall Notice and accompanying Response Form should be included. Questions regarding this notification can be sent to Smiths Medical via email at [email protected].
Nationwide distribution to AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, PR. International distribution
3,952,509 devices total