FDA Recall Terminated

Spacelabs Healthcare elance Vital Signs Monitor, Model 93300, with Option S. It is a patient monitor with arrhythmia detection or alarms.

Recall: Z-0188-2014 · Initiated October 17, 2013

Recall

Recall Number
Z-0188-2014
Event Number
66726
Firm
Spacelabs Healthcare, Llc
FEI Number
3010157426
Product Code
MHX
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
October 17, 2013
Posted
November 8, 2013
Terminated
February 25, 2014
Address
35301 Se Center St, Snoqualmie, WA, 98065-9216

Description

Spacelabs Healthcare elance Vital Signs Monitor, Model 93300, with Option S. It is a patient monitor with arrhythmia detection or alarms.

Reason

Spacelabs Healthcare elance Vital Signs Monitor with Option S, model 93330, is recalled because the monitor will not permit connection of any of the Spacelabs TruLink line of SpO2 sensors.

Action

Spacelabs Healthcare sent the "URGENT -MEDICAL DEVICE CORRECTION Spacelabs Healthcare elance Vital Signs Monitor with Option S" letter, dated 01 November 2013, to the US consignee on November 1, 2013. Spacelabs Healthcare will send the customer letter (translated as necessary) to all international subsidiaries and distributors of record via email on November 6, 2013. USERS ARE ADVISED THAT: -They may not be able to monitor their patient's oxygen saturation because the monitor will not allow the TruLink SpO2 sensor to be connected. The clinician will be required to find an alternate method to monitor oxygen saturation. -They will be contacted at the earliest possible date to schedule a convenient time and Spacelabs will correct all of your facility's affected monitors at no cost. Customer in the US can call Spacelabs Healthcare, Inc. of Snoqualmie, Washington at 800-522-7025. If you have any questions about this corrective action program, please contact Spacelabs Healthcare at +86 512 87178304 for Technical Support.

Distribution

Worldwide Distribution-Distributed in Puerto Rico and the countries of France, Indonesia, Korea, Nicaragua, Philippines, and Taiwan.

Quantity

89 units total (1 in US and 88 International)