FDA Recall Terminated

Safire Duo Ablation Catheter MediGuide Enabled, REF A700240 and A700238, Rx only, Sterile EO. Intended for creating endocardial lesions during cardiac ablation procedures (mapping, stimulation and ablation) for the treatment of typical atrial flutter.

Recall: Z-0187-2014 · Initiated October 11, 2013

Recall

Recall Number
Z-0187-2014
Event Number
66574
Firm
St. Jude Medical
FEI Number
3005334138
Product Code
OAD
Status
Terminated
Root Cause
Process change control
Initiated
October 11, 2013
Posted
November 8, 2013
Terminated
June 9, 2014
Address
5050 Nathan Ln N, Plymouth, MN, 55442-3209

Description

Safire Duo Ablation Catheter MediGuide Enabled, REF A700240 and A700238, Rx only, Sterile EO. Intended for creating endocardial lesions during cardiac ablation procedures (mapping, stimulation and ablation) for the treatment of typical atrial flutter.

Reason

St. Jude Medical Cardiovascular and Ablation Technologies Division is performing a voluntary notification regarding selected batched of Safire Duo Ablation Catheter MediGuide Enabled. These units have a cable connector configuration that is different than intended.

Action

Consignees were sent a St. Jude Medical "Urgent Medical Device Notice" dated 11 October 2013. The letter was addressed to Cath lab manger / Risk Manger. The letter described the problem and the product involved in the recall. The letter stated that it is acceptable to continue using these devices as long as the product is used in accordance with the IFU.

Distribution

Nationwide Distribution including the states of KY, OH, TX, MA, and NC.

Quantity

46